Temperature variation during apicectomy with Er:YAG laser

Photomed Laser Surg. 2012 Aug;30(8):425-8. doi: 10.1089/pho.2012.3256. Epub 2012 Jun 25.

Abstract

Objective: The purpose of this in vitro study was to evaluate the generated temperature of the Er:YAG laser, with three different pulse durations for apicectomy, compared with tungsten bur and surgical saw.

Background data: Apicectomy is an endodontic surgery performed to remove the root apex and curette adjacent periapical tissue because of lesions of the apical area that are not healing properly.

Methods: Sixty single-rooted extracted human teeth were resected by three cutting methods: tungsten bur, surgical saw, and Er:YAG laser irradiation with three different pulse durations; pulse duration 50 μs, pulse duration 100 μs, and pulse duration 300 μs. Teflon-insulated, type K thermocouples were used to measure temperature changes during the apicectomy process. Data were analyzed using the general linear models procedure of the SPSS statistical software program.

Results: Although there was no statistically significant difference for the mean values of temperature changes at 1 mm away to the cutting site of teeth, there was statistically significant difference among groups for the mean values of temperature changes at 3 mm away to the cutting site of teeth. Additionally, there was statistically significant difference among groups for the total time required for apicectomy.

Conclusions: The laser irradiation with pulse duration 50 μs appears to have the lowest temperature rise and the shortest time required for apicectomy of the three pulse durations. However, Er:YAG laser for apicectomy in all pulse durations could be used safely for resection in endodontics in the presence of sufficient water.

MeSH terms

  • Apicoectomy / instrumentation*
  • Dental Instruments
  • Humans
  • In Vitro Techniques
  • Laser Therapy / instrumentation*
  • Lasers, Solid-State*
  • Temperature*
  • Tungsten

Substances

  • Tungsten