Electrical stimulation with Pt electrodes. VIII. Electrochemically safe charge injection limits with 0.2 ms pulses
- PMID: 2276759
- DOI: 10.1109/10.61038
Electrical stimulation with Pt electrodes. VIII. Electrochemically safe charge injection limits with 0.2 ms pulses
Abstract
The charge injection limits of a Pt electrode using 0.2 ms charge balanced, biphasic current pulses ranged from 50 to 150 microC/cm2 geometric if the potential excursions of the electrode are kept below those at which H2 or O2 is produced. These charge densities are three to ten times smaller than the currently accepted value based on earlier experiments in which the reversible surface reactions were fully utilized and the pulse widths were longer.
Similar articles
-
In vitro comparison of the charge-injection limits of activated iridium oxide (AIROF) and platinum-iridium microelectrodes.IEEE Trans Biomed Eng. 2005 Sep;52(9):1612-4. doi: 10.1109/TBME.2005.851503. IEEE Trans Biomed Eng. 2005. PMID: 16189975
-
Offset prediction for charge-balanced stimulus waveforms.J Neural Eng. 2011 Aug;8(4):046032. doi: 10.1088/1741-2560/8/4/046032. Epub 2011 Jul 13. J Neural Eng. 2011. PMID: 21753229
-
Potential-biased, asymmetric waveforms for charge-injection with activated iridium oxide (AIROF) neural stimulation electrodes.IEEE Trans Biomed Eng. 2006 Feb;53(2):327-32. doi: 10.1109/TBME.2005.862572. IEEE Trans Biomed Eng. 2006. PMID: 16485762
-
Neural stimulation and recording electrodes.Annu Rev Biomed Eng. 2008;10:275-309. doi: 10.1146/annurev.bioeng.10.061807.160518. Annu Rev Biomed Eng. 2008. PMID: 18429704 Review.
-
Electrical stimulation of excitable tissue: design of efficacious and safe protocols.J Neurosci Methods. 2005 Feb 15;141(2):171-98. doi: 10.1016/j.jneumeth.2004.10.020. J Neurosci Methods. 2005. PMID: 15661300 Review.
Cited by
-
Simulation of intra-orbital optic nerve electrical stimulation.Med Biol Eng Comput. 2005 Sep;43(5):608-17. doi: 10.1007/BF02351034. Med Biol Eng Comput. 2005. PMID: 16411633
-
Direct Growth of Carbon Nanotubes on New High-Density 3D Pyramid-Shaped Microelectrode Arrays for Brain-Machine Interfaces.Micromachines (Basel). 2016 Sep 8;7(9):163. doi: 10.3390/mi7090163. Micromachines (Basel). 2016. PMID: 30404335 Free PMC article.
-
Contractile Skeletal Muscle Cells Cultured with a Conducting Soft Wire for Effective, Selective Stimulation.Sci Rep. 2018 Feb 2;8(1):2253. doi: 10.1038/s41598-018-20729-y. Sci Rep. 2018. PMID: 29396483 Free PMC article.
-
Retinal prosthesis.IEEE Trans Biomed Eng. 2014 May;61(5):1412-24. doi: 10.1109/TBME.2014.2314733. Epub 2014 Apr 2. IEEE Trans Biomed Eng. 2014. PMID: 24710817 Free PMC article. Review.
-
Chronic electrical stimulation with a suprachoroidal retinal prosthesis: a preclinical safety and efficacy study.PLoS One. 2014 May 22;9(5):e97182. doi: 10.1371/journal.pone.0097182. eCollection 2014. PLoS One. 2014. PMID: 24853376 Free PMC article.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources