Troubleshooting in transatrial tricuspid valve-in-valve implantation

Ann Thorac Surg. 2012 Oct;94(4):1349-52. doi: 10.1016/j.athoracsur.2012.02.084.

Abstract

The established treatment for degenerated stenotic tricuspid bioprostheses is reoperation. Recently, transcatheter tricuspid valve implantation has been reported as an alternative option. This case report describes a complex transcatheter tricuspid valve implantation in a degenerated Medtronic intact 31 mm bioprosthesis. Implantation of a 26 mm Edwards Sapien valve failed, subsequent transcatheter implantation of a 29 mm Edwards Sapien valve was successful.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Bioprosthesis*
  • Cardiac Catheterization / methods*
  • Diagnosis, Differential
  • Echocardiography
  • Female
  • Heart Valve Prosthesis*
  • Humans
  • Prosthesis Design
  • Prosthesis Failure
  • Reoperation / methods*
  • Tomography, X-Ray Computed
  • Tricuspid Valve / surgery*
  • Tricuspid Valve Stenosis / diagnosis