Performance of seven rapid radiographic processing solutions

Oral Surg Oral Med Oral Pathol. 1990 Mar;69(3):382-7. doi: 10.1016/0030-4220(90)90306-d.

Abstract

This study evaluated the performance of seven rapid radiographic processing products. A conventional automatic processing solution was evaluated as a standard of comparison. The products were analyzed with respect to image quality and speed. The results indicate that rapid manual processing generally involves some compromise of image quality as well as slower speeds with an accompanying increase in patient radiation exposure. Accordingly, there seems to be little motivation for the use of rapid manual processing solutions unless there is a compelling rationale in terms of time or convenience.

Publication types

  • Comparative Study

MeSH terms

  • Emulsions
  • Evaluation Studies as Topic
  • Radiography, Dental / methods*
  • Reference Standards
  • Solutions
  • Time Factors
  • X-Ray Film