Application of breast tomosynthesis in screening: incremental effect on mammography acquisition and reading time

Br J Radiol. 2012 Dec;85(1020):e1174-8. doi: 10.1259/bjr/19385909.

Abstract

Objective: The aim of this study was to supplement the paucity of information available on logistical aspects of the application of three-dimensional (3D) mammography in breast screening.

Methods: We prospectively examined the effect on radiographers' and radiologists' workload of implementing 3D mammography in screening by comparing image acquisition time and screen-reading time for two-dimensional (2D) mammography with that of combined 2D+3D mammography. Radiologists' accuracy was also calculated.

Results: Average acquisition time (measured from start of first-view breast positioning to compression release at completion of last view) for seven radiographers, based on 20 screening examinations, was longer for 2D+3D (4 min 3 s; range 3 min 53 s-4 min 18 s) than 2D mammography (3 min 13 s; range 3 min 0 s-3 min 26 s; p<0.01). Average radiologists' reading time per screening examination (three radiologists reading case-mix of 100 screens: 10 cancers, 90 controls) was longer for 2D+3D (77 s; range 60-90 s) than for 2D mammography (33 s; range 25-46 s; p<0.01). 2D+3D screen-reading was associated with detection of more cancers and with substantially fewer recalls than 2D mammography alone.

Conclusion: Relative to standard 2D mammography, combined 2D+3D mammography prolongs image acquisition time and screen-reading time (at initial implementation), and appears to be associated with improved screening accuracy.

Advances in knowledge: These findings provide relevant information to guide larger trials of integrated 3D mammography (2D+3D) and its potential implementation into screening practice.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Breast Neoplasms / diagnostic imaging*
  • Case-Control Studies
  • Clinical Competence / standards*
  • Early Detection of Cancer / methods*
  • Female
  • Humans
  • Imaging, Three-Dimensional / methods
  • Imaging, Three-Dimensional / standards*
  • Mammography / methods
  • Mammography / standards*
  • Prospective Studies
  • Radiology / standards*
  • Time Factors
  • Workload*