Performance of a novel BD stem cell enumeration kit on two flow cytometry systems

Int J Lab Hematol. 2013 Aug;35(4):393-9. doi: 10.1111/ijlh.12029. Epub 2012 Nov 28.

Abstract

Introduction: Hematopoietic stem cell transplantation, which requires accurate enumeration of stem cells, is routinely used in clinical settings. Flow cytometry provides a qualitative and quantitative assessment of CD34⁺ cells. Precision, linearity, and stability of the novel BD™ Stem Cell Enumeration (SCE) Kit were evaluated on two flow cytometry platforms using a modified ISHAGE gating strategy and including a viability dye for data acquisition and analysis.

Methods: Precision and linearity were evaluated on BD FACSCanto™ II and BD FACSCalibur™ systems. Stability was evaluated on the BD FACSCanto II system. Precision was tested using both high and low controls. Linearity was evaluated using dilutions from CD34⁺ cell pools, while stability was evaluated using fresh leukapheresis specimens.

Results: Both systems showed precision with limited variability in absolute counts and percentages of viable CD34⁺ cells. The linearity range of viable CD34⁺ cells in both systems was established at 0-1000 cells/μL, showing a linear relationship (R² = 0.99). Stability of CD34⁺ cells in mobilized leukapheresis samples was confirmed up to 24 h after collection and up to 60 min after the end of stain/lyse procedures.

Conclusion: The BD SCE Kit on both flow cytometry systems shows consistent and reproducible results.

Keywords: Engraftment; flow cytometry; progenitor cells; stem cells; transplantation.

MeSH terms

  • Antigens, CD34 / metabolism
  • Biomarkers / metabolism
  • Calibration
  • Cell Count / standards
  • Cell Survival
  • Flow Cytometry / standards*
  • Hematopoietic Stem Cells / cytology*
  • Hematopoietic Stem Cells / metabolism
  • Humans
  • Leukapheresis
  • Reagent Kits, Diagnostic / standards*
  • Reproducibility of Results

Substances

  • Antigens, CD34
  • Biomarkers
  • Reagent Kits, Diagnostic