The effect of filtering on the two-global-flash mfERG: identifying the optimal range of frequency for detecting glaucomatous retinal dysfunction
- PMID: 23224265
- DOI: 10.1007/s10633-012-9364-8
The effect of filtering on the two-global-flash mfERG: identifying the optimal range of frequency for detecting glaucomatous retinal dysfunction
Abstract
Purpose: To study the effects of filtering bandwidth on the two-global-flash multifocal electroretinogram (mfERG) responses in primary open-angle glaucoma (POAG) compared with control subjects.
Methods: A two-global-flash mfERG (VERIS 6.06™, FMS III) was recorded in 20 healthy subjects and 22 POAG patients with a band-pass filter (BPF) of 1-300 Hz (103 Hexagons, M-sequence stimulus: Lmax 100 cd/m(2), Lmin < 1 cd/m(2), global flash: 200 cd/m(2)). The root-mean-square average of the central 10° was calculated. Three response epochs were analysed: the response to the focal flash, at 15-45 ms (DC), and the following two components induced by the effects of the preceding focal flash on the response to the global flashes at 45-75 ms (IC1) and at 75-105 ms (IC2). The following BPF settings were analysed: 1-300 Hz, 3-300 Hz, 10-300 Hz, 100-300 Hz, 200-300 Hz, 1-10 Hz, 1-100 Hz and 1-200 Hz.
Results: Filtering at 1-300 Hz showed significantly lower responses in POAG than in control subjects (p < 0.001) for all epochs analysed. At 1-100 Hz, this also held true even though the difference between the groups became smaller. At 1-10 Hz, responses were extremely small and did not differ between POAG and control (p > 0.5). This would suggest a filter setting of 10-300 Hz for mfERG recordings in POAG. However, when a filter setting of 10-300 Hz was compared to 1-300 Hz, with a filter setting of 10-300 Hz, the DC in POAG differed more (p < 0.0001) from normal than with 1-300 Hz (p = 0.0002). For IC1 and IC2, the stronger difference between POAG and control was found with 1-300 Hz (p < 0.0001) rather than with 10-300 Hz (p < 0.0001 and p = 0.0005, respectively). For the 'oscillatory potentials' at 100-300 Hz, POAG and control differed significantly in IC1 and IC2 (p < 0.05), but not in DC (p = 0.8). However, filtering at 200-300 Hz did not show a difference between POAG and control (p > 0.5). Thus, we applied a filter setting of 1-200 Hz, which seemed to be most sensitive in detecting glaucomatous retinal dysfunction (p < 0.0001).
Conclusions: A filter setting of 1-200 Hz appears most sensitive to detect glaucomatous damage if using a two-global-flash mfERG: using a band-pass filter a with lower low-frequency cut-off, containing the 10 Hz component, may be especially important in the small induced components that show glaucomatous damage most sensitively. High frequencies of 100-300 Hz also contain information that differentiates glaucoma from normal and thus should be included in the analysis.
Similar articles
-
The 2-global flash mfERG in glaucoma: attempting to increase sensitivity by reducing the focal flash luminance and changing filter settings.Doc Ophthalmol. 2013 Feb;126(1):57-67. doi: 10.1007/s10633-012-9360-z. Epub 2012 Nov 20. Doc Ophthalmol. 2013. PMID: 23179289
-
Impact of a Digital Power Line Filter in the 2-Global-Flash Multifocal Electroretinogram of Glaucoma Patients Compared to Controls.Curr Eye Res. 2016;41(1):70-8. doi: 10.3109/02713683.2014.1002043. Epub 2015 Jan 22. Curr Eye Res. 2016. PMID: 25612055
-
The 'two global flash' mfERG in high and normal tension primary open-angle glaucoma.Doc Ophthalmol. 2007 Jan;114(1):9-19. doi: 10.1007/s10633-006-9033-x. Epub 2006 Dec 8. Doc Ophthalmol. 2007. PMID: 17160421
-
Applications of the multifocal electroretinogram in the detection of glaucoma.Clin Exp Optom. 2011 May;94(3):247-58. doi: 10.1111/j.1444-0938.2010.00571.x. Epub 2011 Feb 16. Clin Exp Optom. 2011. PMID: 21323730 Review.
-
The clinical applications of multifocal electroretinography: a systematic review.Surv Ophthalmol. 2007 Jan-Feb;52(1):61-96. doi: 10.1016/j.survophthal.2006.10.005. Surv Ophthalmol. 2007. PMID: 17212991 Review.
Cited by
-
Novel Machine-Learning Based Framework Using Electroretinography Data for the Detection of Early-Stage Glaucoma.Front Neurosci. 2022 May 4;16:869137. doi: 10.3389/fnins.2022.869137. eCollection 2022. Front Neurosci. 2022. PMID: 35600610 Free PMC article.
-
Ganglion cell layer segmentation and the two-flash multifocal electroretinogram improve structure function analysis in early glaucoma.Graefes Arch Clin Exp Ophthalmol. 2017 Oct;255(10):1991-2000. doi: 10.1007/s00417-017-3722-x. Epub 2017 Aug 4. Graefes Arch Clin Exp Ophthalmol. 2017. PMID: 28779363 Free PMC article.
-
Wavelet decomposition analysis in the two-flash multifocal ERG in early glaucoma: a comparison to ganglion cell analysis and visual field.Doc Ophthalmol. 2017 Aug;135(1):29-42. doi: 10.1007/s10633-017-9593-y. Epub 2017 Jun 7. Doc Ophthalmol. 2017. PMID: 28593391 Free PMC article.
-
Continuous-wavelet-transform analysis of the multifocal ERG waveform in glaucoma diagnosis.Med Biol Eng Comput. 2015 Sep;53(9):771-80. doi: 10.1007/s11517-015-1287-6. Epub 2015 Apr 8. Med Biol Eng Comput. 2015. PMID: 25850982
References
Publication types
MeSH terms