High-sensitivity anastigmatic imaging needle for optical coherence tomography

Opt Lett. 2012 Dec 15;37(24):5247-9. doi: 10.1364/OL.37.005247.

Abstract

We present a high-optical-quality imaging needle for optical coherence tomography (OCT) that achieves sensitivity and resolution comparable to conventional free-space OCT sample arms. The side-viewing needle design utilizes total internal reflection from an angle-polished fiber tip, encased in a glass microcapillary. Fusion of the capillary to the fiber provides a robust, optical-quality output window. The needle's focusing optics are based on an astigmatism-free design, which exploits the "focal shift" phenomenon for focused Gaussian beams to achieve equal working distances (WDs) for both axes. We present a fabricated needle with a WD ratio of 0.98 for imaging in an aqueous environment. Our needle achieves the highest sensitivity of currently reported OCT imaging needles (112 dB), and we demonstrate its performance by superficial imaging of human skin and 3D volumetric imaging within a biological sample.