Assessment of changes in cardiac index and fluid responsiveness: a comparison of Nexfin and transpulmonary thermodilution

Acta Anaesthesiol Scand. 2013 Jul;57(6):704-12. doi: 10.1111/aas.12108. Epub 2013 Mar 25.

Abstract

Background: The Nexfin device uses non-invasive photoplethysmography to monitor cardiac output and respiratory variations in pulse pressure and stroke volume. The aim of this study was to compare rapid changes in cardiac index after fluid challenge between Nexfin and bolus transpulmonary thermodilution and the ability to predict fluid responsiveness of dynamic indices given by Nexfin.

Methods: Simultaneous comparative cardiac index were collected from transpulmonary thermodilution and Nexfin before and after fluid challenge in 45 patients following conventional cardiac surgery. Correlations, Bland-Altman analyses and percentage errors were calculated. Pulse pressure variations and stroke volume variations before fluid challenge were collected to assess their discrimination in predicting fluid responsiveness.

Results: Eight (18%) patients were excluded. A weak positive relationship was found between rapid changes in cardiac index after fluid challenge given by both technologies (n = 37, r = 0.39, P = 0.019). Bias, precision and limits of agreements were 0.20 l/min/m(2) (95% confidence interval (CI) 0.02-0.40), 0.57 l/min/m(2) and ± 1.12 l/min/m(2) before fluid challenge, and 0.01 l/min/m(2) (95% CI -0.24 to 0.26), 0.74 l/min/m(2) and ± 1.45 l/min/m(2) after fluid challenge. Percentage errors between Nexfin and transpulmonary thermodilution were 55% and 58% before and after fluid challenge, respectively. Pulse pressure variations and stroke volume variations given by Nexfin were not discriminant to predict fluid responsiveness: areas under receiver operating characteristics curves 0.57 (95% CI 0.40-0.73) and 0.50 (0.33-0.67), respectively.

Conclusions: The Nexfin cannot be used to measure rapid changes in cardiac index following fluid challenge and to predict fluid responsiveness after cardiac surgery.

Publication types

  • Controlled Clinical Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Area Under Curve
  • Blood Pressure*
  • Body Surface Area
  • Cardiac Output*
  • Cardiac Surgical Procedures
  • Cardiopulmonary Bypass
  • Critical Care / methods*
  • Endpoint Determination
  • Female
  • Fluid Therapy*
  • Humans
  • Male
  • Middle Aged
  • Photoplethysmography / instrumentation*
  • Postoperative Care / instrumentation
  • Postoperative Care / methods*
  • Predictive Value of Tests
  • ROC Curve
  • Sensitivity and Specificity
  • Stroke Volume
  • Thermodilution / methods*
  • Treatment Outcome