Incidence of atrial fibrillation detected by implantable loop recorders in unexplained stroke

Affiliations

Incidence of atrial fibrillation detected by implantable loop recorders in unexplained stroke

Paul E Cotter et al. Neurology. .

Abstract

Objectives: The usefulness of the implantable loop recorder (ILR) with improved atrial fibrillation (AF) detection capability (Reveal XT) and the factors associated with AF in the setting of unexplained stroke were investigated.

Methods: A cohort study is reported of 51 patients in whom ILRs were implanted for the investigation of ischemic stroke for which no cause had been found (cryptogenic) following appropriate vascular and cardiac imaging and at least 24 hours of cardiac rhythm monitoring.

Results: The patients were aged from 17 to 73 (median 52) years. Of the 30 patients with a shunt investigation, 22 had a patent foramen ovale (73.3%; 95% confidence interval [CI] 56.5%-90.1%). AF was identified in 13 (25.5%; 95% CI 13.1%-37.9%) cases. AF was associated with increasing age (p = 0.018), interatrial conduction block (p = 0.02), left atrial volume (p = 0.025), and the occurrence of atrial premature contractions on preceding external monitoring (p = 0.004). The median (range) of monitoring prior to AF detection was 48 (0-154) days.

Conclusion: In patients with unexplained stroke, AF was detected by ILR in 25.5%. Predictors of AF were identified, which may help to target investigations. ILRs may have a central role in the future in the investigation of patients with unexplained stroke.

PubMed Disclaimer

Figures

Figure
Figure. Bar graph of time to atrial fibrillation (AF) detection in cases where AF was detected
Dashed lines represent median and interquartile range. *Case 3 had a delay of 0 days (AF on postimplantation device check).

Comment in

Similar articles

Cited by

References

    1. Marini C, De Santis F, Sacco S, et al. Contribution of atrial fibrillation to incidence and outcome of ischemic stroke: results from a population-based study. Stroke 2005;36:1115–1119 - PubMed
    1. Cotter PE, Belham M, Martin PJ. Stroke in younger patients: the heart of the matter. J Neurol 2010;257:1777–1787 - PubMed
    1. Seet RC, Friedman PA, Rabinstein AA. Prolonged rhythm monitoring for the detection of occult paroxysmal atrial fibrillation in ischemic stroke of unknown cause. Circulation 2011;124:477–486 - PubMed
    1. Ziegler PD, Koehler JL, Mehra R. Comparison of continuous versus intermittent monitoring of atrial arrhythmias. Heart Rhythm 2006;3:1445–1452 - PubMed
    1. Liao J, Khalid Z, Scallan C, Morillo C, O'Donnell M. Noninvasive cardiac monitoring for detecting paroxysmal atrial fibrillation or flutter after acute ischemic stroke: a systematic review. Stroke 2007;38:2935–2940 - PubMed