A pulsatile blood vessel system for a femoral arterial access clinical simulation model
- PMID: 23688608
- DOI: 10.1016/j.medengphy.2013.04.010
A pulsatile blood vessel system for a femoral arterial access clinical simulation model
Abstract
The model-based, rapid-prototyping-enabled design and manufacture of a pulsatile blood vessel (PBV) for high-fidelity mannequin-based clinical simulations is presented. The PBV presented here is a pressurized, flexible tube with alternating fluid pressure created by a pump to mimic the behavior of a human vessel in response to pulsatile pressure. The use of PBVs is important for the fidelity of a clinical simulator that requires residents to palpate and/or access the vessel. In this study, a PBV is presented which features the integration of 3D modeling using patient-specific computed tomography (CT) data, mold fabrication using rapid-prototyping, and finite element method for estimating the required pumping pressure to generate the same level of force (about 1.5 N) experienced by the user through palpation. The relationship between this palpation force and the vessel pressure is studied using two strategies: finite element analysis (FEA) and experiments in a femoral arterial access simulator with a pump, artificial vessel, and surrounding phantom tissue. The experimental results show a discrepancy of 8.7% from the FEA-predicted value. Qualitative validation is done by exposing and surveying 19 interventional cardiology residents at four major educational institutions to the simulator for accuracy of its feel. The overall survey results are positive.
Keywords: Arterial access; Clinical simulator; Finite element analysis.
Copyright © 2013 IPEM. Published by Elsevier Ltd. All rights reserved.
Similar articles
-
Pulsatile arterial wall-blood flow interaction with wall pre-stress computed using an inverse algorithm.Biomed Eng Online. 2015;14 Suppl 1(Suppl 1):S18. doi: 10.1186/1475-925X-14-S1-S18. Epub 2015 Jan 9. Biomed Eng Online. 2015. PMID: 25603022 Free PMC article.
-
A robust 3D finite element simulation of human proximal femur progressive fracture under stance load with experimental validation.Ann Biomed Eng. 2013 Dec;41(12):2515-27. doi: 10.1007/s10439-013-0864-9. Epub 2013 Jul 18. Ann Biomed Eng. 2013. PMID: 23864338
-
Coupled fluid-structure interaction hemodynamics in a zero-pressure state corrected arterial geometry.J Biomech. 2011 Sep 2;44(13):2453-60. doi: 10.1016/j.jbiomech.2011.06.024. Epub 2011 Jul 16. J Biomech. 2011. PMID: 21762918
-
Efficient 3D finite element analysis of dental restorative procedures using micro-CT data.Dent Mater. 2007 May;23(5):539-48. doi: 10.1016/j.dental.2006.03.013. Epub 2006 May 26. Dent Mater. 2007. PMID: 16730058
-
A one-dimensional finite element method for simulation-based medical planning for cardiovascular disease.Comput Methods Biomech Biomed Engin. 2002 Jun;5(3):195-206. doi: 10.1080/10255840290010670. Comput Methods Biomech Biomed Engin. 2002. PMID: 12186712
Cited by
-
Examining the Effect of Haptic Factors for Vascular Palpation Skill Assessment Using an Affordable Simulator.IEEE Open J Eng Med Biol. 2020;1:228-234. doi: 10.1109/ojemb.2020.3017156. Epub 2020 Aug 17. IEEE Open J Eng Med Biol. 2020. PMID: 33681817 Free PMC article.
-
A combination of experimental and finite element analyses of needle-tissue interaction to compute the stresses and deformations during injection at different angles.J Clin Monit Comput. 2016 Dec;30(6):965-975. doi: 10.1007/s10877-015-9801-9. Epub 2015 Oct 29. J Clin Monit Comput. 2016. PMID: 26515741
-
Applications of three-dimensional printing technology in the cardiovascular field.Intern Emerg Med. 2015 Oct;10(7):769-80. doi: 10.1007/s11739-015-1282-9. Epub 2015 Jul 29. Intern Emerg Med. 2015. PMID: 26220346
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources