Adiabatic inversion pulses for myocardial T1 mapping

Magn Reson Med. 2014 Apr;71(4):1428-34. doi: 10.1002/mrm.24793. Epub 2013 May 30.

Abstract

Purpose: To evaluate the error in T1 estimates using inversion-recovery-based T1 mapping due to imperfect inversion and to perform a systematic study of adiabatic inversion pulse designs in order to maximize inversion efficiency for values of transverse relaxation (T2) in the myocardium subject to a peak power constraint.

Methods: The inversion factor for hyperbolic secant and tangent/hyperbolic tangent adiabatic full passage waveforms was calculated using Bloch equations. A brute-force search was conducted for design parameters: pulse duration, frequency range, shape parameters, and peak amplitude. A design was selected that maximized the inversion factor over a specified range of amplitude and off-resonance and validated using phantom measurements. Empirical correction for imperfect inversion was performed.

Results: The tangent/hyperbolic tangent adiabatic pulse was found to outperform hyperbolic secant designs and achieve an inversion factor of 0.96 within ±150 Hz over 25% amplitude range with 14.7 µT peak amplitude. T1 mapping errors of the selected design due to imperfect inversion was ∼4% and could be corrected to <1%.

Conclusions: Nonideal inversion leads to significant errors in inversion-recovery-based T1 mapping. The inversion efficiency of adiabatic pulses is sensitive to transverse relaxation. The tangent/hyperbolic tangent design achieved the best performance subject to the peak amplitude constraint.

Keywords: MOLLI; MRI; T1 mapping; adiabatic inversion.

Publication types

  • Research Support, N.I.H., Intramural

MeSH terms

  • Algorithms*
  • Heart / anatomy & histology*
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / methods*
  • Phantoms, Imaging
  • Reproducibility of Results
  • Sensitivity and Specificity