Long-term survival of the uncemented Balgrist total hip replacement cup

Int Orthop. 2013 Aug;37(8):1449-56. doi: 10.1007/s00264-013-1946-x. Epub 2013 Jun 8.

Abstract

Purpose: The purpose of this study was to evaluate the long-term clinical and radiographic outcomes of the Balgrist total hip replacement cup.

Methods: We report the results of a retrospective review of 195 total hip prostheses with Balgrist cup implanted in 178 patients. There were 59 men and 119 women with a mean age at surgery of 52.3 years (range, 19-72). Cox regression was used to determine the influence of independent variables on the survivorship of the prosthesis.

Results: There were 117 (66 %) of 178 patients, and 131 (67 %) of 195 hips available for assessment at a mean follow-up of 17.4 years. A total of 21 patients (22 hips) were lost from follow-up (12 %). A total of 25 patients (27 hips) required some type of re-operation during the follow-up period. Eight of 27 re-operations (30 %) were directly caused by the Balgrist cup failure due to fatigue fracture of the metallic shell (n = 5; 19 %), polyethylene wear (n = 2; 7 %) and aseptic loosening (n = 1; 4 %). The mean polyethylene wear rate was 0.068 mm/year (range, 0.008-0.230; SD = 0.043). The risk of re-operation for any reason was decreased with dysplastic hip as a primary diagnosis (hazard ratio; HR = 0.250; 95 % CI 0.086-0.725) and with greater age at the time of surgery (HR = 0.923; 95 % CI 0.880-0.969). The 19-year survivorship with re-operation for any reason was 75.6 % (95 % CI 67.0-84.2). The 19-year survivorship with re-operation for the Balgrist cup failure was 90.5 % (95 % CI 83.4-97.6).

Conclusions: The Balgrist cup continues to provide excellent clinical and radiological outcomes. This is associated at least in part with a low polyethylene wear rate. The main reason for the Balgrist cup failure is fatigue fracture of the metallic shell.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Arthroplasty, Replacement, Hip / instrumentation*
  • Bone Cements
  • Cohort Studies
  • Female
  • Follow-Up Studies
  • Hip Dislocation / surgery*
  • Hip Joint / diagnostic imaging
  • Hip Joint / surgery*
  • Hip Prosthesis*
  • Humans
  • Incidence
  • Kaplan-Meier Estimate
  • Longitudinal Studies
  • Male
  • Middle Aged
  • Osteoarthritis, Hip / surgery*
  • Polyethylene
  • Proportional Hazards Models
  • Prosthesis Design*
  • Radiography
  • Reoperation / instrumentation
  • Reoperation / statistics & numerical data
  • Retrospective Studies
  • Treatment Outcome

Substances

  • Bone Cements
  • Polyethylene