Diffusion tensor imaging of the median nerve at 3.0 T using different MR scanners: agreement of FA and ADC measurements
- PMID: 23768611
- DOI: 10.1016/j.ejrad.2013.05.011
Diffusion tensor imaging of the median nerve at 3.0 T using different MR scanners: agreement of FA and ADC measurements
Abstract
Objective: To assess the agreement of fractional anisotropy (FA) and apparent diffusion coefficient (ADC) values of the median nerve on 3.0 T MR scanners from different vendors.
Materials and methods: IRB approved study including 16 healthy volunteers (9 women; mean age 30.6 ± 5.3 years). Diffusion tensor imaging (DTI) of the dominant wrist was performed on three 3.0 T MR scanners (GE, Siemens, Philips) using similar imaging protocols and vendor-proprietary hard- and software. Intra-, inter-reader and inter-vendor agreements were assessed.
Results: ICCs for intra-/inter-reader agreements ranged from 0.843-0.970/0.846-0.956 for FA, and 0.840-0.940/0.726-0.929 for ADC, respectively. ANOVA analysis identified significant differences for FA/ADC measurements among vendors (p < 0.001/p < 0.01, respectively). Overall mean values for FA were 0.63 (SD ± 0.1) and 0.999 × 10(-3)mm(2)/s (SD ± 0.134 × 10(-3)) for ADC. A significant negative measurement bias was found for FA values from the GE scanner (-0.05 and -0.07) and for ADC values from the Siemens scanner (-0.053 and -0.063 × 10(-3)mm(2)/s) as compared to the remainder vendors
Conclusion: FA and ADC values of the median nerve obtained on different 3.0 T MR scanners differ significantly, but are in comparison to the standard deviation of absolute values small enough to not have an impact on larger group studies or when substantial diffusion changes can be expected. However, caution is warranted in an individual patient when interpreting diffusion values from different scanner acquisitions.
Keywords: Diffusion tensor imaging; Magnetic resonance imaging; Median nerve; Peripheral neuropathy.
Copyright © 2013 Elsevier Ireland Ltd. All rights reserved.
Similar articles
-
Assessment of median nerve with MR neurography by using diffusion-tensor imaging: normative and pathologic diffusion values.Radiology. 2012 Oct;265(1):194-203. doi: 10.1148/radiol.12111403. Epub 2012 Jul 19. Radiology. 2012. PMID: 22820733
-
Apparent diffusion coefficient reproducibility of the pancreas measured at different MR scanners using diffusion-weighted imaging.J Magn Reson Imaging. 2014 Dec;40(6):1375-81. doi: 10.1002/jmri.24492. Epub 2013 Nov 12. J Magn Reson Imaging. 2014. PMID: 24222019
-
Comparison of the diagnostic performances of diffusion parameters in diffusion weighted imaging and diffusion tensor imaging of breast lesions.Eur J Radiol. 2013 Dec;82(12):e801-6. doi: 10.1016/j.ejrad.2013.09.001. Epub 2013 Sep 13. Eur J Radiol. 2013. PMID: 24099642
-
Assessment of the diagnostic value of diffusion tensor imaging in patients with spinal cord compression: a meta-analysis.Braz J Med Biol Res. 2016 Jan;49(1):e4769. doi: 10.1590/1414-431X20154769. Epub 2015 Nov 27. Braz J Med Biol Res. 2016. PMID: 26628393 Free PMC article. Review.
-
Diffusion tensor imaging of peripheral nerves.Skeletal Radiol. 2010 Nov;39(11):1073-9. doi: 10.1007/s00256-010-0974-5. Epub 2010 Jul 1. Skeletal Radiol. 2010. PMID: 20593175 Review.
Cited by
-
Magnetic resonance neurography and diffusion tensor imaging of the sciatic nerve in hereditary transthyretin amyloidosis polyneuropathy.J Neurol. 2023 Oct;270(10):4827-4840. doi: 10.1007/s00415-023-11813-z. Epub 2023 Jun 17. J Neurol. 2023. PMID: 37329346 Free PMC article.
-
Correlation between diffusion tensor indices and fascicular morphometric parameters of peripheral nerve.Front Physiol. 2023 Feb 23;14:1070227. doi: 10.3389/fphys.2023.1070227. eCollection 2023. Front Physiol. 2023. PMID: 36909220 Free PMC article.
-
Diffusion-weighted MR imaging of musculoskeletal tissues: incremental role over conventional MR imaging in bone, soft tissue, and nerve lesions.BJR Open. 2022 Mar 18;4(1):20210077. doi: 10.1259/bjro.20210077. eCollection 2022. BJR Open. 2022. PMID: 36452057 Free PMC article. Review.
-
Quantitative MR-Neurography at 3.0T: Inter-Scanner Reproducibility.Front Neurosci. 2022 Feb 16;16:817316. doi: 10.3389/fnins.2022.817316. eCollection 2022. Front Neurosci. 2022. PMID: 35250457 Free PMC article.
-
Meta-analysis of the normal diffusion tensor imaging values of the median nerve and how they change in carpal tunnel syndrome.Sci Rep. 2021 Oct 22;11(1):20935. doi: 10.1038/s41598-021-00353-z. Sci Rep. 2021. PMID: 34686721 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources