Biodegradable magnesium-based screw clinically equivalent to titanium screw in hallux valgus surgery: short term results of the first prospective, randomized, controlled clinical pilot study
- PMID: 23819489
- PMCID: PMC3702514
- DOI: 10.1186/1475-925X-12-62
Biodegradable magnesium-based screw clinically equivalent to titanium screw in hallux valgus surgery: short term results of the first prospective, randomized, controlled clinical pilot study
Erratum in
-
Correction to: Biodegradable magnesium-based screw clinically equivalent to titanium screw in hallux valgus surgery: short term results of the first prospective, randomized, controlled clinical pilot study.Biomed Eng Online. 2020 Nov 23;19(1):86. doi: 10.1186/s12938-020-00818-8. Biomed Eng Online. 2020. PMID: 33225926 Free PMC article.
Abstract
Purpose: Nondegradable steel-and titanium-based implants are commonly used in orthopedic surgery. Although they provide maximal stability, they are also associated with interference on imaging modalities, may induce stress shielding, and additional explantation procedures may be necessary. Alternatively, degradable polymer implants are mechanically weaker and induce foreign body reactions. Degradable magnesium-based stents are currently being investigated in clinical trials for use in cardiovascular medicine. The magnesium alloy MgYREZr demonstrates good biocompatibility and osteoconductive properties. The aim of this prospective, randomized, clinical pilot trial was to determine if magnesium-based MgYREZr screws are equivalent to standard titanium screws for fixation during chevron osteotomy in patients with a mild hallux valgus.
Methods: Patients (n=26) were randomly assigned to undergo osteosynthesis using either titanium or degradable magnesium-based implants of the same design. The 6 month follow-up period included clinical, laboratory, and radiographic assessments.
Results: No significant differences were found in terms of the American Orthopaedic Foot and Ankle Society (AOFAS) score for hallux, visual analog scale for pain assessment, or range of motion (ROM) of the first metatarsophalangeal joint (MTPJ). No foreign body reactions, osteolysis, or systemic inflammatory reactions were detected. The groups were not significantly different in terms of radiographic or laboratory results.
Conclusion: The radiographic and clinical results of this prospective controlled study demonstrate that degradable magnesium-based screws are equivalent to titanium screws for the treatment of mild hallux valgus deformities.
Figures
Similar articles
-
Hallux valgus correction utilising a modified short scarf osteotomy with a magnesium biodegradable or titanium compression screws - a comparative study of clinical outcomes.BMC Musculoskelet Disord. 2019 Jul 19;20(1):334. doi: 10.1186/s12891-019-2717-7. BMC Musculoskelet Disord. 2019. PMID: 31319832 Free PMC article.
-
Magnesium-based bioabsorbable screw fixation for hallux valgus surgery - A suitable alternative to metallic implants.Foot Ankle Surg. 2019 Dec;25(6):727-732. doi: 10.1016/j.fas.2018.09.001. Epub 2018 Sep 22. Foot Ankle Surg. 2019. PMID: 30321917
-
Comparison of Bioabsorbable Magnesium versus Titanium Screw Fixation for Modified Distal Chevron Osteotomy in Hallux Valgus.Biomed Res Int. 2018 Nov 19;2018:5242806. doi: 10.1155/2018/5242806. eCollection 2018. Biomed Res Int. 2018. PMID: 30581858 Free PMC article.
-
Comparison between bioabsorbable magnesium and titanium compression screws for hallux valgus treated with distal metatarsal osteotomies: A meta-analysis.Jt Dis Relat Surg. 2023 May 12;34(2):289-297. doi: 10.52312/jdrs.2023.1026. Jt Dis Relat Surg. 2023. PMID: 37462631 Free PMC article. Review.
-
Modified Lindgren-Turan Osteotomy for Hallux Valgus Deformity - a Review of 60 Cases.Acta Chir Orthop Traumatol Cech. 2018;85(5):325-330. Acta Chir Orthop Traumatol Cech. 2018. PMID: 30383528 Review. English.
Cited by
-
Tuning biodegradability, bone-bonding capacity, and wear resistance of zinc-30% magnesium intermetallic alloy for use in load-bearing bone applications.Sci Rep. 2024 Jan 29;14(1):2425. doi: 10.1038/s41598-024-52648-6. Sci Rep. 2024. PMID: 38287092 Free PMC article.
-
Magnesium Alloys in Orthopedics: A Systematic Review on Approaches, Coatings and Strategies to Improve Biocompatibility, Osteogenic Properties and Osteointegration Capabilities.Int J Mol Sci. 2023 Dec 24;25(1):282. doi: 10.3390/ijms25010282. Int J Mol Sci. 2023. PMID: 38203453 Free PMC article. Review.
-
Results of the Use of Bioabsorbable Magnesium Screws for Surgical Treatment of Mason Type II Radial Head Fractures.Clin Orthop Surg. 2023 Dec;15(6):1013-1021. doi: 10.4055/cios23074. Epub 2023 Oct 25. Clin Orthop Surg. 2023. PMID: 38045577 Free PMC article.
-
Magnesium-based alloys with adapted interfaces for bone implants and tissue engineering.Regen Biomater. 2023 Nov 1;10:rbad095. doi: 10.1093/rb/rbad095. eCollection 2023. Regen Biomater. 2023. PMID: 38020233 Free PMC article. Review.
-
Comparison of Resorbable and Non-Resorbable Osteosynthesis Material in Hallux Surgery: A Systematic Review.Life (Basel). 2023 Oct 5;13(10):2018. doi: 10.3390/life13102018. Life (Basel). 2023. PMID: 37895399 Free PMC article. Review.
References
-
- Waizy H, Seitz JM, Reifenrath J, Weizbauer A, Bach FW, Meyer-Lindenberg A, Denkena B, Windhagen H. Biodegradable magnesium implants for orthopedic applications. J Mater Sci. 2013;48:39–50. doi: 10.1007/s10853-012-6572-2. - DOI
-
- Bostman OM, Paivarinta U, Partio E, Manninen M, Vasenius J, Majola A, Rokkanen P. The tissue-implant interface during degradation of absorbable polyglycolide fracture fixation screws in the rabbit femur. Clin Orthop Relat Res. 1992;285:263–272. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources