Radiographically measured total knee wear is constant and predicts failure

J Arthroplasty. 2013 Sep;28(8):1338-44. doi: 10.1016/j.arth.2013.04.047. Epub 2013 Jul 10.

Abstract

Knee wear is commonly measured with the most recent radiograph based on the assumption that wear progresses at a constant rate. Changing patient activity or in vivo polyethylene deterioration are examples of factors that could cause wear rates to change over time. Using six or more radiographs on each of 251 knees over a mean 10-year follow-up, we determined the pattern of polyethylene wear. 92% of knees had linear wear with a mean wear rate of 0.09 ± 0.12 mm/yr. Ten knees (4%) had late accelerated wear. Knees with accelerated wear and those with a wear rate of 0.15 mm/yr or greater had lower survivorship rates. We conclude that wear is linear with rare exceptions and that higher wear is correlated with failure.

Keywords: medial tibial polyethylene thickness loss; primary total knee arthroplasty; revision; serial radiographic analysis; survivorship; wear pattern.