Effect of target distance on accommodative amplitude measured using the minus lens technique

Clin Exp Optom. 2014 Jan;97(1):62-5. doi: 10.1111/cxo.12090. Epub 2013 Jul 28.

Abstract

Background: The aim was to investigate the effect on the measured amplitude of accommodation and repeatability of using the minus lens technique with the target at distance or near.

Methods: Forty-three students (average age: 21.17 ± 1.50 years, 35 female) had their amplitude of accommodation measured with minus lenses on top of their distance correction in a trial frame with the target at far (6.0 m) or near (0.4 m). The minus lens power was gradually added with steps of 0.25 D. Measurements were taken on two occasions at each distance, which were separated by a time interval of at least 24 hours.

Results: The measured amplitude at six metres was significantly lower than that with the target at 40 cm, by 1.56 ± 1.17 D (p < 0.001) and this varied between individuals (r = 0.716, intraclass correlation coefficient = 0.439). With either target distance, repeated measurement was highly correlated (r > 0.9) but the agreement was better at 6.0 m (±0.74 D) than at 40 cm (± 0.92 D).

Conclusion: The measurements of the amplitude of accommodation with the minus lens technique using targets at far or near are not comparable and the difference between the target distances may provide clinically relevant information.

Keywords: accommodation amplitude; agreement; minus lens method; repeatability.

Publication types

  • Clinical Trial
  • Validation Study

MeSH terms

  • Accommodation, Ocular*
  • Adolescent
  • Female
  • Humans
  • Male
  • Optometry / instrumentation
  • Optometry / methods*
  • Prospective Studies
  • Refractive Errors / diagnosis*
  • Reproducibility of Results
  • Retinoscopy / methods*
  • Retinoscopy / standards*
  • Young Adult