Detection and identification of occult HBV in blood donors in Taiwan using a commercial, multiplex, multi-dye nucleic acid amplification technology screening test

Vox Sang. 2014 Feb;106(2):103-10. doi: 10.1111/vox.12075. Epub 2013 Aug 2.

Abstract

Background: The ability of a new generation commercial, multiplex, multi-dye test from Roche, the cobas TaqScreen MPX test, version 2.0, to detect and identify occult HBV infections was evaluated using routine donor samples from Kaohsiung Blood Bank, Taiwan.

Study design and methods: A total of 5973 samples were tested by nucleic acid amplification technology (NAT); 5898 in pools of six, 66 in pools of less than six and nine samples individually. NAT-reactive samples were retested with alternative NAT tests, and follow-up samples from the donors were tested individually by NAT and for all the HBV serological markers.

Results: Eight NAT-only-reactive donors were identified, and follow-up samples were obtained from six of the donors. The results indicated that all eight donors had an occult HBV infection with viral loads <12 IU/ml.

Conclusion: The cobas(®) TaqScreen MPX test, version 2.0, has an advantage over the current Roche blood screening test, the cobas TaqScreen MPX test, for screening donations in countries with a high prevalence of occult HBV infections since the uncertainty associated with identifying samples with very low viremia is removed by the ability of the test to identify the viral target in samples that are reactive with the cobas TaqScreen MPX test, version 2.0.

Keywords: NAT test; blood screening; multi-dye; occult HBV.

Publication types

  • Evaluation Study

MeSH terms

  • Blood Donors*
  • DNA, Viral / blood
  • Hepatitis B virus / genetics
  • Hepatitis B virus / isolation & purification*
  • Humans
  • Mass Screening / methods
  • Nucleic Acid Amplification Techniques / methods*
  • Serologic Tests
  • Taiwan