The inferior vena cava clip. The percutaneous approach

Invest Radiol. 1990 Aug;25(8):871-5. doi: 10.1097/00004424-199008000-00001.

Abstract

Pulmonary embolism in high-risk patients may be minimized by surgical inferior vena cava (IVC) clipping or by the insertion of caval filters. A percutaneous clipping technique was developed that narrows the cava while allowing caval patency. The caval clip is inserted through a percutaneous translumbar approach under fluoroscopic control. Nine dogs underwent percutaneous translumbar caval clip placement without complications. Three of four dogs, followed-up for 5 to 19 weeks by angiography and caval pressure measurements, showed caval patency. This technique eliminates the risks of surgical IVC clip placement and risks from the insertion of intravascular foreign bodies such as filters.

MeSH terms

  • Angiography
  • Animals
  • Constriction
  • Dogs
  • Female
  • Fluoroscopy
  • Male
  • Prostheses and Implants*
  • Pulmonary Embolism / prevention & control*
  • Radiology, Interventional
  • Stainless Steel
  • Vena Cava, Inferior*

Substances

  • Stainless Steel