Complication with intraosseous access: scandinavian users' experience
- PMID: 24106537
- PMCID: PMC3789903
- DOI: 10.5811/westjem.2013.1.12000
Complication with intraosseous access: scandinavian users' experience
Abstract
Introduction: Intraosseous access (IO) is indicated if vascular access cannot be quickly established during resuscitation. Complication rates are estimated to be low, based on small patient series, model or cadaver studies, and case reports. However, user experience with IO use in real-life emergency situations might differ from the results in the controlled environment of model studies and small patient series. We performed a survey of IO use in real-life emergency situations to assess users' experiences of complications.
Methods: An online questionnaire was sent to Scandinavian emergency physicians, anesthesiologists and pediatricians.
Results: 1,802 clinical cases of IO use was reported by n=386 responders. Commonly reported complications with establishing IO access were patient discomfort/pain (7.1%), difficulties with penetration of periosteum with IO needle (10.3%), difficulties with aspiration of bone marrow (12.3%), and bended/broken needle (4.0%). When using an established IO access the reported complications were difficulties with injection fluid and drugs after IO insertion (7.4%), slow infusion (despite use of pressure bag) (8.8%), displacement after insertion (8.5%), and extravasation (3.7%). Compartment syndrome and osteomyelitis occurred in 0.6% and 0.4% of cases respectively.
Conclusion: In users' recollection of real-life IO use, perceived complications were more frequent than usually reported from model studies. The perceived difficulties with using IO could affect the willingness of medical staff to use IO. Therefore, user experience should be addressed both in education of how to use, and research and development of IOs.
Similar articles
-
Efficacy of the EZ-IO needle driver for out-of-hospital intraosseous access--a preliminary, observational, multicenter study.Scand J Trauma Resusc Emerg Med. 2011 Oct 26;19:65. doi: 10.1186/1757-7241-19-65. Scand J Trauma Resusc Emerg Med. 2011. PMID: 22029625 Free PMC article.
-
Emergency intraosseous access in a helicopter emergency medical service: a retrospective study.Scand J Trauma Resusc Emerg Med. 2010 Oct 7;18:52. doi: 10.1186/1757-7241-18-52. Scand J Trauma Resusc Emerg Med. 2010. PMID: 20929544 Free PMC article.
-
Comparison of the Fluid Resuscitation Rate with and without External Pressure Using Two Intraosseous Infusion Systems for Adult Emergencies, the CITRIN (Comparison of InTRaosseous infusion systems in emergency medicINe)-Study.PLoS One. 2015 Dec 2;10(12):e0143726. doi: 10.1371/journal.pone.0143726. eCollection 2015. PLoS One. 2015. PMID: 26630579 Free PMC article.
-
Intraosseous drug administration in children and adults during cardiopulmonary resuscitation.Ann Pharmacother. 2007 Oct;41(10):1679-86. doi: 10.1345/aph.1K168. Epub 2007 Aug 14. Ann Pharmacother. 2007. PMID: 17698894 Review.
-
Tibial Intraosseous Insertion in Pediatric Emergency Care: A Review Based upon Postmortem Computed Tomography.Prehosp Emerg Care. 2020 Sep-Oct;24(5):665-671. doi: 10.1080/10903127.2019.1698682. Epub 2020 Jan 7. Prehosp Emerg Care. 2020. PMID: 31774707 Review.
Cited by
-
Compartment Syndrome Resulting From Improper Intraosseous Cannulation: A Case Report.Cureus. 2023 Dec 9;15(12):e50248. doi: 10.7759/cureus.50248. eCollection 2023 Dec. Cureus. 2023. PMID: 38196424 Free PMC article.
-
Finding the most suitable puncture site for intraosseous access in term and preterm neonates: an ultrasound-based anatomical pilot study.Eur J Pediatr. 2023 Jul;182(7):3083-3091. doi: 10.1007/s00431-023-04972-8. Epub 2023 Apr 19. Eur J Pediatr. 2023. PMID: 37074459 Free PMC article.
-
Intraosseous Tibial Resuscitation After a Total Knee Arthroplasty Leading to Osteonecrosis and Loosening of the Tibial Component.Arthroplast Today. 2023 Jan 14;19:101088. doi: 10.1016/j.artd.2022.101088. eCollection 2023 Feb. Arthroplast Today. 2023. PMID: 36691465 Free PMC article.
-
Intraosseous access in neonates is feasible and safe - An analysis of a prospective nationwide surveillance study in Germany.Front Pediatr. 2022 Jul 26;10:952632. doi: 10.3389/fped.2022.952632. eCollection 2022. Front Pediatr. 2022. PMID: 35958173 Free PMC article.
-
Intraosseous fluid resuscitation causes systemic fat emboli in a porcine hemorrhagic shock model.Scand J Trauma Resusc Emerg Med. 2021 Dec 20;29(1):172. doi: 10.1186/s13049-021-00986-z. Scand J Trauma Resusc Emerg Med. 2021. PMID: 34930433 Free PMC article.
References
-
- Luck RP, Haines C, Mull CC. Intraosseous access. J Emerg Med. 2010;39:468–475. - PubMed
-
- Gerritse BM, Scheffer GJ, Draaisma JM. Prehospital intraosseus access with the bone injection gun by a helicopter-transported emergency medical team. J Trauma. 2009;66:1739–1741. - PubMed
-
- Fiorito BA, Mirza F, Doran TM, et al. Intraosseous access in the setting of pediatric critical care transport. Pediatr Crit Care Med. 2005;6:50–53. - PubMed
-
- Jun H, Haruyama AZ, Chang KS, et al. Comparison of a new screw-tipped intraosseous needle versus a standard bone marrow aspiration needle for infusion. Am J Emerg Med. 2000;18:135–139. - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources