Cardiovasc Interv Ther. 2014 Jul;29(3):247-51. doi: 10.1007/s12928-013-0220-x. Epub 2013 Oct 25.

Abstract

A 67-year-old woman was admitted with an acute coronary syndrome. Angiographic examination revealed diffuse severe stenosis of the left circumflex artery. A Pressure Wire Certus (St. Jude Medical, St. Paul, MN, USA) was passed to the second post-lateral (PL) branch. To prevent side-branch occlusion, a SION wire (Asahi Intecc Co., Aichi, Japan) was left in the first PL branch, and a SION blue wire (Asahi Intecc) was placed in the second obtuse marginal branch. We implanted an everolimus-eluting stent (PROMUS Element 2.5 × 24 mm, Boston Scientific, Natick, MA, USA) in the culprit lesion. After retrieving the protection wire in first PL branch with resistance, we performed post-dilatation. However, the intravascular ultrasound images showed that the proximal portion of the implanted stent had elongated approximately 2 mm to the left main trunk (LMT), although the position of the distal edge of the stent was unchanged. We decided to additionally place a stent from the ostium of the LMT to the proximal left anterior descending coronary artery, and a biolimus-eluting stent (NOBORI 3.0 × 18 mm, Terumo Co., Tokyo, Japan) was implanted successfully. Longitudinal stent elongation might be caused by the small number of links between the hoops of a stent, originally intended to improve deliverability.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Blood Vessel Prosthesis Implantation / adverse effects
  • Blood Vessel Prosthesis Implantation / methods
  • Coronary Stenosis / surgery*
  • Coronary Vessels / diagnostic imaging
  • Coronary Vessels / surgery
  • Female
  • Humans
  • Percutaneous Coronary Intervention / adverse effects
  • Percutaneous Coronary Intervention / methods
  • Radiography
  • Stents / adverse effects*
  • Ultrasonography, Interventional