Stress generation during self-adjusting file movement: minimally invasive instrumentation

J Endod. 2013 Dec;39(12):1572-5. doi: 10.1016/j.joen.2013.07.021. Epub 2013 Aug 31.

Abstract

Introduction: Although nickel-titanium (NiTi) rotary instruments may produce a well- tapered root canal with a low tendency of aberrations, these are generally perceived to have a high fracture risk during use and may produce significant forces on root dentin during instrumentation, which may induce a dentinal defect or crack in the apical part of the root. This study compared mathematically the stress generated by the Self-Adjusting File (ReDent-Nova, Ra'anana, Israel) with conventional rotary instruments during the movement of 3 NiTi endodontic file designs in a curved root canal.

Methods: Stresses were calculated using finite element analysis. Three file designs with tip size ISO #20 were used in this study. Finite element models of ProFile #20/.06 (Dentsply Maillefer, Ballaigues, Switzerland) (a constant tapered shaft), ProTaper Universal F1 (Dentsply Maillefer) (a progressively changing taper shaft), and SAF 1.5 mm (a mesh shaft) were activated within a curved root canal model. The stress generations resulting from the simulated shaping movement were evaluated in the apical root dentin area.

Results: The SAF induced the lowest von Mises stress concentration and the lowest tensile principal stress component in root dentin. The calculated stress values from ProTaper Universal F1 and ProFile #20/.06 were approximately 8 to 10 times bigger than that of the SAF.

Conclusions: Stress levels during shaping and the susceptibility to apical root cracks after shaping vary with instrument design. The design of the SAF may produce minimal stress concentrations in the apical root dentin during shaping of the curved canal, which may increase the chance of preservation of root dentin integrity with a reduced risk of dentinal defects and apical root cracking.

Keywords: Finite element analysis; Self-Adjusting File; nickel-titanium rotary file; root integrity; stress generation; vertical root fracture.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Computer-Aided Design
  • Dental Alloys / chemistry
  • Dental Pulp Cavity / anatomy & histology
  • Dentin / anatomy & histology
  • Equipment Design
  • Finite Element Analysis
  • Humans
  • Imaging, Three-Dimensional / methods
  • Materials Testing
  • Movement
  • Nickel / chemistry
  • Root Canal Preparation / instrumentation*
  • Rotation
  • Stress, Mechanical
  • Surface Properties
  • Tensile Strength
  • Titanium / chemistry
  • Tooth Apex / anatomy & histology
  • X-Ray Microtomography / methods

Substances

  • Dental Alloys
  • titanium nickelide
  • Nickel
  • Titanium