Ultrasensitive electrochemical immunosensors for multiplexed determination using mesoporous platinum nanoparticles as nonenzymatic labels

Anal Chim Acta. 2014 Jan 7:807:44-50. doi: 10.1016/j.aca.2013.11.025. Epub 2013 Nov 23.

Abstract

An ultrasensitive multiplexed immunoassay method was developed at a disposable immunosensor array using mesoporous platinum nanoparticles (M-Pt NPs) as nonenzymatic labels. M-Pt NPs were prepared by ultrasonic method and employed to label the secondary antibody (Ab2) for signal amplification. The immunosensor array was constructed by covalently immobilizing capture antibody (Ab1) on graphene modified screen printed carbon electrodes (SPECs). After the sandwich-type immunoreactions, the M-Pt-Ab2 was bound to immunosensor surface to catalyze the electro-reduction of H2O2 reaction, which produced detectable signals for readout of analytes. Using breast cancer related panel of tumor markers (CA125, CA153 and CEA) as model analytes, this method showed wide linear ranges of over 4 orders of magnitude with the detection limits of 0.002 U mL(-1), 0.001 U mL(-1) and 7.0 pg mL(-1) for CA125, CA153 and CEA, respectively. The disposable immunosensor array possessed excellent clinical value in cancer screening as well as convenient point of care diagnostics.

Keywords: Immunosensor array; Mesoporous platinum nanoparticles; Nonenzymatic; Point of care diagnostics; Screen printed carbon electrodes.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Antibodies, Immobilized / chemistry
  • Antibodies, Immobilized / immunology
  • Biomarkers, Tumor / blood*
  • Biosensing Techniques
  • CA-125 Antigen / blood
  • Carcinoembryonic Antigen / blood
  • Electrochemical Techniques*
  • Electrodes
  • Graphite / chemistry
  • Immunoassay*
  • Metal Nanoparticles / chemistry*
  • Mucin-1 / blood
  • Platinum / chemistry*
  • Porosity

Substances

  • Antibodies, Immobilized
  • Biomarkers, Tumor
  • CA-125 Antigen
  • Carcinoembryonic Antigen
  • Mucin-1
  • Platinum
  • Graphite