On the quantification accuracy, homogeneity, and stability of simultaneous positron emission tomography/magnetic resonance imaging systems

Invest Radiol. 2014 Jun;49(6):373-81. doi: 10.1097/RLI.0000000000000021.

Abstract

Objective: A potential major application of simultaneous avalanche photodiode-based positron emission tomography (PET)/magnetic resonance imaging (MRI) systems are quantitative brain studies for cerebral blood flow measurements in combination with blood-oxygen-level-dependent or perfusion MRI, requiring a high performance for both modalities. Thus, we evaluated PET quantification accuracy and homogeneity for 2 different simultaneous PET/MRI systems (whole-body and brain scanner) compared with those of a state-of-the-art PET detector (PET/computed tomography) using phantom studies. In addition, we investigated the long-term stability of PET and quality of functional MRI measurements of a clinical whole-body PET/MRI scanner.

Materials and methods: Phantom measurements were conducted using spheres filled with [F]-fluoride distributed in a homogeneous cylinder phantom at different positions inside the PET field of view. Recovery values and standard deviations were extracted from resulting PET images. The influence of magnetic resonance-based attenuation correction and that of activity outside the PET field of view on the recovery values of these spheres was evaluated. Furthermore, long-term PET stability of the whole-body PET/MRI system was assessed by evaluating position profiles, energy spectra, count rates, and recovery values from [Ge]-phantom scans. Functional MRI applicability was tested in accordance with the functional Biomedical Information Research Network procedure.

Results: The BrainPET system showed high recovery values (up to 99%) but also increased variability (up to 7.4%). Significant underestimations in PET quantification near activity outside the PET field of view were found (up to 80%). Using magnetic resonance-based attenuation correction led to an underestimation in PET activity of approximately 7%. In distinction, the whole-body PET/MRI system revealed performance similar to the PET/computed tomographic scanner (recovery values up to approximately 60% with a variability of approximately 4%). Long-term stability and fMRI performance of the whole-body PET/MRI scanner showed no degradation compared with stand-alone systems.

Conclusions: Homogeneity and accuracy of avalanche photodiode-based PET detectors is comparable with those of the state-of-the-art detectors based on photomultiplier tubes. However, attenuation correction on PET/MRI systems has to be adapted carefully for quantitative PET measurements. The BrainPET system needs improved scatter correction to perform quantitative brain studies. The whole-body PET/MRI scanner, however, is applicable for quantitative brain studies.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Brain*
  • Humans
  • Magnetic Resonance Imaging / instrumentation*
  • Multimodal Imaging*
  • Phantoms, Imaging
  • Positron-Emission Tomography / instrumentation*
  • Tomography, X-Ray Computed / instrumentation*
  • Whole Body Imaging / instrumentation*