A novel method for experimental characterization of large-angle scattered particles in scanned carbon-ion therapy

Med Phys. 2014 Feb;41(2):021706. doi: 10.1118/1.4860256.

Abstract

Purpose: It is essential to consider large-angle scattered particles in dose calculation models for therapeutic carbon-ion beams. However, it is difficult to measure the small dose contribution from large-angle scattered particles. In this paper, the authors present a novel method to derive the parameters describing large-angle scattered particles from the measured results.

Methods: The authors developed a new parallel-plate ionization chamber consisting of concentric electrodes. Since the sensitive volume of each channel is increased linearly with this type, it is possible to efficiently and easily detect small contributions from the large-angle scattered particles. The parameters describing the large-angle scattered particles were derived from pencil beam dose distribution in water measured with the new ionization chamber. To evaluate the validity of this method, the correction for the field-size dependence of the doses, "predicted-dose scaling factor," was calculated with the new parameters.

Results: The predicted-dose scaling factor calculated with the new parameters was compared with the existing one. The difference between the new correction factor and the existing one was 1.3%. For target volumes of different sizes, the calculated dose distribution with the new parameters was in good agreement with the measured one.

Conclusions: Parameters describing the large-angle scattered particles can be efficiently and rapidly determined using the new ionization chamber. The authors confirmed that the field-size dependence of the doses could be compensated for by the new parameters. This method makes it possible to easily derive the parameters describing the large-angle scattered particles, while maintaining the dose calculation accuracy.

MeSH terms

  • Electrodes
  • Heavy Ion Radiotherapy / methods*
  • Radiometry
  • Radiotherapy Dosage
  • Reproducibility of Results
  • Scattering, Radiation*