Inspiratory Flows and Volumes in Subjects with Non-CF Bronchiectasis Using a New Dry Powder Inhaler Device

Open Respir Med J. 2014 Jan 24:8:8-13. doi: 10.2174/1874306401408010008. eCollection 2014.

Abstract

Introduction: Drug inhalation via a dry-powder inhaler (DPI) is a convenient, time efficient alternative to nebulizers in the treatment of cystic fibrosis (CF) or non-CF bronchiectasis. Efficient drug administration via DPIs depends on the device resistance and adequate (≥45L/min) inspiratory flows and volumes generated by individuals. Drypowder mannitol is delivered using a RS01 breath-actuated device developed by Plastiape, for Pharmaxis. The study aim was to determine in vivo if non-CF bronchiectasis patients' inspiratory flows and volumes are adequate to use the RS01 DPI device.

Materials and methodology: An open, non-interventional study; enrolled 17 subjects with non-CF bronchiectasis, 18 to 80 years, with baseline FEV1 ≥1.0L and ≥50‰ predicted. Inspiratory flows and volumes were measured when subjects inhaled in a controlled manner through the RS01 device in series with a spirometer.

Results: The mean inspiratory volume (IV) of non-CF bronchiectasis subjects was 2.08 ± 0.5L and achieved a mean PIF of 78.6 ± 11.2L/min with the inhaler device.

Conclusion: Use of the RS01 DPI device allowed adequate inspiratory flow and volume for dispersion of dry-powder mannitol in non-CF bronchiectasis patients.

Keywords: Bronchiectasis; clinical trials; dry powder inhaler; high resistance inhaler; peak inspiratory flow.; respiratory function tests.