Comparison of magnetic resonance feature tracking for systolic and diastolic strain and strain rate calculation with spatial modulation of magnetization imaging analysis

J Magn Reson Imaging. 2015 Apr;41(4):1000-12. doi: 10.1002/jmri.24623. Epub 2014 Mar 28.

Abstract

Purpose: To compare cardiovascular magnetic resonance-feature tracking (CMR-FT) with spatial modulation of magnetization (SPAMM) tagged imaging for the calculation of short and long axis Lagrangian strain measures in systole and diastole.

Materials and methods: Healthy controls (n = 35) and patients with dilated cardiomyopathy (n = 10) were identified prospectively and underwent steady-state free precession (SSFP) cine imaging and SPAMM imaging using a gradient-echo sequence. A timed offline analysis of images acquired at identical horizontal long and short axis slice positions was performed using CMR-FT and dynamic tissue-tagging (CIMTag2D). Agreement between strain and strain rate (SR) values calculated using these two different methods was assessed using the Bland-Altman technique.

Results: Across all participants, there was good agreement between CMR-FT and CIMTag for calculation of peak systolic global circumferential strain (-22.7 ± 6.2% vs. -22.5 ± 6.9%, bias 0.2 ± 4.0%) and SR (-1.35 ± 0.42 1/s vs. -1.22 ± 0.42 1/s, bias 0.13 ± 0.33 1/s) and early diastolic global circumferential SR (1.21 ± 0.44 1/s vs. 1.07 ± 0.30 1/s, bias -0.14 ± 0.34 1/s) at the subendocardium. There was satisfactory agreement for derivation of peak systolic global longitudinal strain (-18.1 ± 5.0% vs. -16.7 ± 4.8%, bias 1.3 ± 3.8%) and SR (-1.04 ± 0.29 1/s vs. -0.95 ± 0.32 1/s, bias 0.09 ± 0.26 1/s). The weakest agreement was for early diastolic global longitudinal SR (1.10 ± 0.40 1/s vs. 0.67 ± 0.32 1/s, bias -0.42 ± 0.40 1/s), although the correlation remained significant (r = 0.42, P < 0.01). CMR-FT generated these data over four times quicker than CIMTag.

Conclusion: There is sufficient agreement between systolic and diastolic strain measures calculated using CMR-FT and myocardial tagging for CMR-FT to be considered as a potentially feasible and rapid alternative.

Keywords: cine magnetic resonance imaging; feature tracking; left ventricular function; tagging.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Cardiomyopathy, Dilated / complications
  • Cardiomyopathy, Dilated / physiopathology*
  • Diastole
  • Elastic Modulus
  • Elasticity Imaging Techniques / methods*
  • Heart Ventricles / physiopathology*
  • Humans
  • Image Enhancement / methods
  • Image Interpretation, Computer-Assisted / methods*
  • Magnetic Resonance Imaging, Cine / methods*
  • Middle Aged
  • Pattern Recognition, Automated / methods
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Shear Strength
  • Stress, Mechanical
  • Systole
  • Ventricular Dysfunction / etiology
  • Ventricular Dysfunction / physiopathology*