Carbamylation of albumin is a cause for discrepancies between albumin assays

Clin Chim Acta. 2014 Jul 1:434:6-10. doi: 10.1016/j.cca.2014.03.035. Epub 2014 Apr 5.

Abstract

Background: Several investigators have reported discrepancies between the bromocresol-purple (BCP), bromocresol-green (BCG) and immunonephelometric (INP) assays in dialysis patients. This study compared the abovementioned assays and investigated whether hemodialysis itself or carbamylation of albumin is the cause for this discrepancy.

Methods: Samples obtained from hemodialysis patients were analyzed by BCP, BCG and INP. Furthermore, albumin was carbamylated in vitro using isocyanate. Isocyanate converts lysine to homocitrulline.

Results: No differences were observed between samples of the pre- and post-hemodialysis groups for BCP. In the control group, BCG averaged 6 g/L higher than INP. BCP did not statistically deviate from INP. In the dialysis group BCG averaged 5 g/L higher when compared to INP, whereas BCP averaged 2 g/L lower. BCP was affected by carbamylation of albumin. BCG and INP measurements were affected to a much lesser extent. Homocitrulline content of hydrolysates was increased in both the carbamylated albumin as well as in the dialysis population.

Conclusion: In a hemodialysis population albumin concentrations are not consistently estimated by both BCG and BCP methods. Relative to INP measurements BCG overestimates the albumin concentration (4-10 g/L), whereas BCP leads to an underestimation (0-4 g/L). Carbamylation of albumin is the main attributor to the discrepancy found with BCP.

Keywords: Albumin; BCG; BCP; Bromocresol; Carbamylation; Hemodialysis.