A high-sensitivity method for the measurement of 222Rn based on liquid scintillation counting of polycarbonate powder

Radiat Prot Dosimetry. 2014 Jul;160(1-3):188-91. doi: 10.1093/rpd/ncu075. Epub 2014 Apr 10.

Abstract

This work describes a technique for the measurement of 222Rn by absorption in polycarbonate (PC) powder and liquid scintillation counting (LSC). The work is an improvement of the recently proposed method for 222Rn measurements by LSC of exposed PC grains. It is demonstrated that the use of PC powder as a 222Rn sampler improves 13.6 times the sampling efficiency and leads to 6.5 times smaller minimmal detectable activity concentrations (MDAC) compared with the PC grains used so far. For a 40-h exposure of 7.4-g PC powder to 222Rn in air, the MDAC with a RackBeta 1219 LS counter is 62 Bq m(-3) (assuming a 8-h sample counting time and 24-h background time). For the same conditions the estimated 222Rn MDAC with a Quantulus 1220 LS counter is 20 Bq m(-3). The proposed technique is suitable for radon in air and radon in soil-gas measurements.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Air Pollutants, Radioactive / analysis*
  • Humans
  • Polycarboxylate Cement / chemistry*
  • Powders / chemistry*
  • Radiation Monitoring / instrumentation*
  • Radiation Monitoring / methods
  • Radon / analysis*
  • Scintillation Counting / instrumentation
  • Scintillation Counting / methods*
  • Soil Pollutants, Radioactive / analysis*

Substances

  • Air Pollutants, Radioactive
  • Polycarboxylate Cement
  • Powders
  • Soil Pollutants, Radioactive
  • polycarbonate
  • Radon