Self-designed posterior atlas polyaxial lateral mass screw-plate fixation for unstable atlas fracture

Spine J. 2014 Dec 1;14(12):2892-6. doi: 10.1016/j.spinee.2014.04.020. Epub 2014 Apr 24.

Abstract

Background context: Most atlas fractures can be effectively treated nonoperatively with external immobilization unless there is an injury to the transverse atlantal ligament. Surgical stabilization is most commonly achieved using a posterior approach with fixation of C1-C2 or C0-C2, but these treatments usually result in loss of the normal motion of the C1-C2 and C0-C1 joints.

Purpose: To clinically validate feasibility, safety, and value of open reduction and fixation using an atlas polyaxial lateral mass screw-plate construct in unstable atlas fractures.

Study design: Retrospective review of patients who sustained unstable atlas fractures treated with polyaxial lateral mass screw-plate construct.

Patient sample: Twenty-two patients with unstable atlas fractures who underwent posterior atlas polyaxial lateral mass screw-plate fixation were analyzed.

Outcome measures: Visual analog scale, neurologic status, and radiographs for fusion.

Methods: From January 2011 to September 2012, 22 patients with unstable atlas fractures were treated with this technique. Patients' charts and radiographs were reviewed. Bone fusion, internal fixation placement, and integrity of spinal cord and vertebral arteries were assessed via intraoperative and follow-up imaging. Neurologic function, range of motion, and pain levels were assessed clinically on follow-up.

Results: All patients were followed up from 12 to 32 months, with an average of 22.5±18.0 months. A total of 22 plates were placed, and all 44 screws were inserted into the atlas lateral masses. The mean duration of the procedure was 86 minutes, and the average estimated blood loss was 120 mL. Computed tomography scans 9 months after surgery confirmed that fusion was achieved in all cases. There was no screw or plate loosening or breakage in any patient. All patients had well-preserved range of motion. No vascular or neurologic complication was noted, and all patients had a good clinical outcome.

Conclusions: An open reduction and posterior internal fixation with atlas polyaxial lateral mass screw-plate is a safe and effective surgical option in the treatment of unstable atlas fractures. This technique can provide immediate reduction and preserve C1-C2 motion.

Keywords: Atlas fractures; Internal fixation; Open reduction; Posterior approach; Transverse atlantal ligament; Unstable fracture.

Publication types

  • Clinical Trial
  • Validation Study

MeSH terms

  • Adult
  • Aged
  • Atlanto-Axial Joint / injuries
  • Atlanto-Axial Joint / surgery
  • Atlanto-Occipital Joint / injuries
  • Atlanto-Occipital Joint / surgery
  • Bone Plates*
  • Bone Screws*
  • Cervical Atlas / injuries
  • Cervical Atlas / surgery*
  • Female
  • Fracture Fixation, Internal / instrumentation*
  • Fracture Fixation, Internal / methods
  • Humans
  • Male
  • Middle Aged
  • Spinal Fractures / surgery*
  • Spinal Fusion / instrumentation
  • Spinal Fusion / methods*