Test-retest reliability of automated whole body and compartmental muscle volume measurements on a wide bore 3T MR system

Eur Radiol. 2014 Sep;24(9):2279-91. doi: 10.1007/s00330-014-3226-6. Epub 2014 May 29.

Abstract

Purpose: To measure the test-retest reproducibility of an automated system for quantifying whole body and compartmental muscle volumes using wide bore 3 T MRI.

Materials and methods: Thirty volunteers stratified by body mass index underwent whole body 3 T MRI, two-point Dixon sequences, on two separate occasions. Water-fat separation was performed, with automated segmentation of whole body, torso, upper and lower leg volumes, and manually segmented lower leg muscle volumes.

Results: Mean automated total body muscle volume was 19·32 L (SD9·1) and 19·28 L (SD9·12) for first and second acquisitions (Intraclass correlation coefficient (ICC) = 1·0, 95% level of agreement -0·32-0·2 L). ICC for all automated test-retest muscle volumes were almost perfect (0·99-1·0) with 95% levels of agreement 1.8-6.6% of mean volume. Automated muscle volume measurements correlate closely with manual quantification (right lower leg: manual 1·68 L (2SD0·6) compared to automated 1·64 L (2SD 0·6), left lower leg: manual 1·69 L (2SD 0·64) compared to automated 1·63 L (SD0·61), correlation coefficients for automated and manual segmentation were 0·94-0·96).

Conclusion: Fully automated whole body and compartmental muscle volume quantification can be achieved rapidly on a 3 T wide bore system with very low margins of error, excellent test-retest reliability and excellent correlation to manual segmentation in the lower leg.

Key points: Sarcopaenia is an important reversible complication of a number of diseases. Manual quantification of muscle volume is time-consuming and expensive. Muscles can be imaged using in and out of phase MRI. Automated atlas-based segmentation can identify muscle groups. Automated muscle volume segmentation is reproducible and can replace manual measurements.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Body Fluids / physiology*
  • Female
  • Healthy Volunteers
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / instrumentation*
  • Male
  • Middle Aged
  • Muscle, Skeletal / anatomy & histology*
  • Organ Size
  • Prospective Studies
  • Reproducibility of Results
  • Whole Body Imaging / methods*