Imaging dose in variable pitch body perfusion CT scans: an analysis using TG111 formalism
- PMID: 24877823
- DOI: 10.1118/1.4876377
Imaging dose in variable pitch body perfusion CT scans: an analysis using TG111 formalism
Abstract
Purpose: To investigate the variation of imaging dose with tube potential in variable pitch body CT perfusion (CTp) protocols using the TG111 dosimetric formalism.
Methods: TG111 recommendations were followed in choosing the phantom, dosimetric equipment, and methodology. Specifically, equilibrium doses (D(eq)) were measured centrally and peripherally in a long PMMA phantom. Reference planar average equilibrium doses were determined for each tube potential, for a reference set of exposure parameters (collimation, pitch, filtration) on a Siemens Definition CT scanner. These reference values were utilized to predict the imaging dose during perfusion scans using interpretations of the TG111 formalism. As a gold reference, the midscan average planar perfusion doses (D(CTp)) were obtained directly from central and peripheral D(eq) measurements for body CTp scans (144 and 271 mm) using variable pitch acquisition. Measurement-based D(CTp) values obtained using a thimble chamber were compared to the TG111-predicted values, and to CTDI(vol) reported at the console.
Results: Reference planar average equilibrium dose values measured for reference uniform pitch helical scans were consistently higher than console-reported or measured values for CTDI(vol). The measurement-based perfusion dose D(CTp) was predicted accurately by the reported CTDI(vol) for the 144 mm scan. The 271 mm scans delivered systematically larger dose than reported. The TG111-based dose estimates were proven to be conservative, as they were systematically higher than both the measured and the reported imaging doses.
Conclusions: Upon successful implementation of TG111 formalism, standard imaging dose was measured for a body CTp protocol using the variable pitch helical acquisition. The TG111 formalism is not directly applicable to this type of acquisition. Measurement of dose for all variable pitch protocols is strongly suggested.
Similar articles
-
A database for estimating organ dose for coronary angiography and brain perfusion CT scans for arbitrary spectra and angular tube current modulation.Med Phys. 2012 Sep;39(9):5336-46. doi: 10.1118/1.4739243. Med Phys. 2012. PMID: 22957601
-
A comprehensive Monte Carlo study of CT dose metrics proposed by the AAPM Reports 111 and 200.Med Phys. 2022 Jan;49(1):201-218. doi: 10.1002/mp.15306. Epub 2021 Dec 13. Med Phys. 2022. PMID: 34800303
-
Influence of phantom diameter, kVp and scan mode upon computed tomography dose index.Med Phys. 2003 Mar;30(3):395-402. doi: 10.1118/1.1543149. Med Phys. 2003. PMID: 12674240
-
Basis and current state of computed tomography perfusion imaging: a review.Phys Med Biol. 2022 Sep 7;67(18). doi: 10.1088/1361-6560/ac8717. Phys Med Biol. 2022. PMID: 35926503 Review.
-
CT perfusion in oncology: how to do it.Cancer Imaging. 2010 Feb 11;10(1):8-19. doi: 10.1102/1470-7330.2010.0001. Cancer Imaging. 2010. PMID: 20159664 Free PMC article. Review.
Cited by
-
Pilot study of combined FDG-PET and dynamic contrast-enhanced CT of locally advanced cervical carcinoma before and during concurrent chemoradiotherapy suggests association between changes in tumor blood volume and treatment response.Cancer Med. 2018 Aug;7(8):3642-3651. doi: 10.1002/cam4.1632. Epub 2018 Jul 2. Cancer Med. 2018. PMID: 29963760 Free PMC article. Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials