Image quality and radiation dose of lower extremity CT angiography at 70 kVp on an integrated circuit detector dual-source computed tomography

Acta Radiol. 2015 Jun;56(6):659-65. doi: 10.1177/0284185114535391. Epub 2014 Jun 11.

Abstract

Background: Despite the well-established requirement for radiation dose reduction there are few studies examining the potential for lower extremity CT angiography (CTA) at 70 kVp.

Purpose: To compare the image quality and radiation dose of lower extremity CTA at 70 kVp using a dual-source CT system with an integrated circuit detector to similar studies at 120 kVp.

Material and methods: A total of 62 patients underwent lower extremity CTA. Thirty-one patients were examined at 70 kVp using a second generation dual-source CT with an integrated circuit detector (70 kVp group) and 31 patients were evaluated at 120 kVp using a first generation dual-source CT (120 kVp group). The attenuation and image noise were measured and signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. Two radiologists assessed image quality. Radiation dose was compared.

Results: The mean attenuation of the 70 kVp group was higher than the 120 kVp group (575 ± 149 Hounsfield units [HU] vs. 258 ± 38 HU, respectively, P < 0.001) as was SNR (44.0 ± 22.0 vs 32.7 ± 13.3, respectively, P = 0.017), CNR (39.7 ± 20.6 vs 26.6 ± 11.7, respectively, P = 0.003) and the mean image quality score (3.7 ± 0.1 vs. 3.2 ± 0.3, respectively, P < 0.001). The inter-observer agreement was good for the 70 kVp group and moderate for the 120 kVp group. The dose-length product was lower in the 70 kVp group (264.5 ± 63.1 mGy × cm vs. 412.4 ± 81.5 mGy × cm, P < 0.001).

Conclusion: Lower extremity CTA at 70 kVp allows for lower radiation dose with higher SNR, CNR, and image quality when compared with standard 120 kVp.

Keywords: Angiography; X-ray computed tomography (CT); low tube voltage; radiation dosage.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Angiography / methods*
  • Feasibility Studies
  • Female
  • Humans
  • Lower Extremity / blood supply*
  • Lower Extremity / diagnostic imaging*
  • Male
  • Middle Aged
  • Radiation Dosage*
  • Retrospective Studies
  • Tomography, X-Ray Computed / methods*
  • Young Adult