Purpose: To evaluate sensitivity, specificity and reproducibility of colour difference plot analysis (CDPA) of 103 hexagon multifocal electroretinogram (mfERG) in detecting established hydroxychloroquine (HCQ) retinal toxicity.
Methods: Twenty-three patients taking HCQ were divided into those with and without retinal toxicity and were compared with a control group without retinal disease and not taking HCQ. CDPA with two masked examiners was performed using age-corrected mfERG responses in the central ring (Rc ; 0-5.5 degrees from fixation) and paracentral ring (Rp ; 5.5-11 degrees from fixation). An abnormal ring was defined as containing any hexagons with a difference in two or more standard deviations from normal (colour blue or black).
Results: Categorical analysis (ring involvement or not) showed Rc had 83% sensitivity and 93% specificity. Rp had 89% sensitivity and 82% specificity. Requiring abnormal hexagons in both Rc and Rp yielded sensitivity and specificity of 83% and 95%, respectively. If required in only one ring, they were 89% and 80%, respectively. In this population, there was complete agreement in identifying toxicity when comparing CDPA using Rp with ring ratio analysis using R5/R4 P1 ring responses (89% sensitivity and 95% specificity). Continuous analysis of CDPA with receiver operating characteristic analysis showed optimized detection (83% sensitivity and 96% specificity) when ≥4 abnormal hexagons were present anywhere within the Rp ring outline. Intergrader agreement and reproducibility were good.
Conclusions: Colour difference plot analysis had sensitivity and specificity that approached that of ring ratio analysis of R5/R4 P₁ responses. Ease of implementation and reproducibility are notable advantages of CDPA.
Keywords: hydroxychloroquine toxicity; multifocal electroretinography; retinal; toxicity; toxicology.
© 2014 Acta Ophthalmologica Scandinavica Foundation. Published by John Wiley & Sons Ltd.