De novo aortic regurgitation after continuous-flow left ventricular assist device implantation
- PMID: 25069685
- DOI: 10.1016/j.athoracsur.2014.05.030
De novo aortic regurgitation after continuous-flow left ventricular assist device implantation
Abstract
Background: Significant aortic regurgitation (AR) after continuous-flow left ventricular assist device (cf-LVAD) placement affects device performance and patient outcomes. This study examined the development of AR and long-term results after implantation of cf-LVADs.
Methods: The study included all patients with no or less than mild AR who underwent HeartMate II (58 [62%]; Thoratec Corp, Pleasanton, CA) or HeartWare (35 [38%]; HeartWare International, Framingham, MA) implantation at our institute from July 2006 to July 2012. Serial echocardiograms were obtained preoperatively, at 1, 3 and 6 months postoperatively, and then at a minimum of 4-month intervals in patients with longer-term support. Kaplan-Meier estimates for freedom from moderate or greater AR were generated. Logistic regression analysis was used to define independent predictors of AR after cf-LVAD implantation.
Results: Median duration of LVAD support was 527 days (25(th), 75(th): 289, 907; range, 60 to 2,433 days). Mild AR developed in 48 patients (51.6%) over a median duration of 126 days, with progression to moderate AR in 13 (14%) over 493 days and to severe AR in 2 (2.1%) over 1,231 days. The incidence of mild or greater AR was 43.1% in HeartMate II vs 65.7% in HeartWare recipients (p = 0.035). Overall freedom from moderate or greater AR was 94.7% ± 2.6% at 1 year, 86.9% ± 4.5% at 2 years, 82.8% ± 5.9% at 3 years, and 31% ± 16.9% at 4 years. Independent predictors of AR were duration of support (odds ratio, 1.002; 95% confidence interval, 1.000 to 1.004; p = 0.017) and a persistently closed aortic valve (odds ratio, 0.193; 95% confidence interval, 0.097 to 0.382; p < 0.001).
Conclusions: AR is associated with longer cf-LVAD support duration and persistent aortic valve closure. Incidence of moderate or greater AR after cf-LVAD implantation increases significantly after 3 years. The clinical implications of these data may warrant consideration of prophylactic aortic valve replacement at the time of cf-LVAD implantation, particularly with expected longer duration of support and in patients with preexisting AR that is more than mild.
Copyright © 2014 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Does postoperative blood pressure influence development of aortic regurgitation following continuous-flow left ventricular assist device implantation?†.Eur J Cardiothorac Surg. 2016 Mar;49(3):788-94. doi: 10.1093/ejcts/ezv221. Epub 2015 Jun 23. Eur J Cardiothorac Surg. 2016. PMID: 26108216
-
Surgical correction of aortic valve insufficiency after left ventricular assist device implantation.J Thorac Cardiovasc Surg. 2013 Nov;146(5):1247-52. doi: 10.1016/j.jtcvs.2013.05.019. Epub 2013 Jul 16. J Thorac Cardiovasc Surg. 2013. PMID: 23870154
-
The development of aortic insufficiency in continuous-flow left ventricular assist device-supported patients.Ann Thorac Surg. 2013 Feb;95(2):493-8. doi: 10.1016/j.athoracsur.2012.09.020. Epub 2012 Dec 13. Ann Thorac Surg. 2013. PMID: 23245444
-
De Novo Aortic Regurgitation After Continuous-Flow Left Ventricular Assist Device Implantation.Ann Thorac Surg. 2017 Aug;104(2):704-711. doi: 10.1016/j.athoracsur.2017.01.114. Epub 2017 May 6. Ann Thorac Surg. 2017. PMID: 28483150 Review.
-
De novo aortic insufficiency during long-term support on a left ventricular assist device: a systematic review and meta-analysis.ASAIO J. 2014 Mar-Apr;60(2):183-8. doi: 10.1097/MAT.0000000000000042. ASAIO J. 2014. PMID: 24399060 Review.
Cited by
-
Aortic valve disorders and left ventricular assist devices.Front Cardiovasc Med. 2023 Feb 23;10:1098348. doi: 10.3389/fcvm.2023.1098348. eCollection 2023. Front Cardiovasc Med. 2023. PMID: 36910539 Free PMC article. Review.
-
Reciprocal interferences of the left ventricular assist device and the aortic valve competence.Front Cardiovasc Med. 2023 Jan 9;9:1094796. doi: 10.3389/fcvm.2022.1094796. eCollection 2022. Front Cardiovasc Med. 2023. PMID: 36698950 Free PMC article. Review.
-
Concomitant or late aortic valve intervention and its efficacy for aortic insufficiency associated with continuous-flow left ventricular assist device implantation.Front Cardiovasc Med. 2022 Nov 15;9:1029984. doi: 10.3389/fcvm.2022.1029984. eCollection 2022. Front Cardiovasc Med. 2022. PMID: 36457799 Free PMC article. Review.
-
Current Role of the Total Artificial Heart in the Management of Advanced Heart Failure.Curr Cardiol Rep. 2019 Nov 22;21(11):142. doi: 10.1007/s11886-019-1242-5. Curr Cardiol Rep. 2019. PMID: 31758343 Review.
-
The angle of the outflow graft to the aorta can affect recirculation due to aortic insufficiency under left ventricular assist device support.J Artif Organs. 2018 Dec;21(4):399-404. doi: 10.1007/s10047-018-1064-z. Epub 2018 Jul 23. J Artif Organs. 2018. PMID: 30039455