Radiation exposure to hand surgeons' hands: a practical comparison of large and mini C-arm fluoroscopy

J Hand Surg Am. 2014 Sep;39(9):1805-9. doi: 10.1016/j.jhsa.2014.06.133. Epub 2014 Jul 31.

Abstract

Purpose: To determine and compare the radiation exposure to surgeons' hands with large and mini C-arm fluoroscopy in a practical, clinically based model.

Methods: Two hand surgeons monitored radiation exposure to their hands with a ring dosimeter over a 14-month period using large and mini C-arm fluoroscopic units. One surgeon performed all cases with a large C-arm unit in a hospital setting, and the other performed all cases with mini C-arms in surgical centers. For each case, fluoroscopic time, the output displayed by the unit, radiation by time, and ring dosimeter absorption were recorded and analyzed.

Results: A total of 160 consecutive cases were reviewed with 71 cases and 89 cases in the large and mini C-arm groups, respectively. The median output displayed by the large C-arm was 0.7 mGy/case, and the median output displayed by the mini C-arm was 10.0 mGy/case. With output as a product of time, the median calculated values were 0.02 mGy/s for the large C-arm group and 0.28 mGy/s for the mini C-arm group. Cumulative ring dosimeter absorption to the surgeons' hands was found to be 380 mrem for 71 cases in the large C-arm group versus 1,000 mrem for 89 cases in the mini C-arm group.

Conclusions: In our model, the use of the mini C-arm resulted in more than a 10-fold increase in the rate of output and approximately double the dosimeter absorption to the surgeon's hand compared with the large C-arm. Although it has been shown that the mini C-arm produces less radiation scatter, in a practical model, it may not be a safer alternative with respect to the surgeon's hands. Based on these findings, we recommend that surgeons be more aware of radiation exposure risk, know their C-arm unit's specifications, and try to minimize radiation exposure.

Type of study/level of evidence: Prognostic II.

Keywords: Fluoroscopy; large C-arm; mini C-arm; radiation; radiation exposure.

Publication types

  • Comparative Study

MeSH terms

  • Fluoroscopy / instrumentation*
  • Hand / radiation effects*
  • Hand / surgery*
  • Humans
  • Occupational Exposure*
  • Prospective Studies
  • Radiation Dosage
  • Radiation Monitoring / instrumentation
  • Radiation Protection
  • Surgeons*