Quantitative determination of total L-carnitine in infant formula, follow-up formula, and raw materials by liquid chromatography with tandem mass spectrometry

Anal Sci. 2014;30(8):839-44. doi: 10.2116/analsci.30.839.

Abstract

We developed a rapid and useful routine screening assay for total L-carnitine content in various infant formulas and materials by liquid chromatography coupled with a tandem mass spectrometric method (LC-MS/MS) and alkaline hydrolysis. For separation of L-carnitine, a multi-mode octadecylsilane (ODS) column was used that contained ODS ligands, anion ligands, and cation ligands to avoid using ion-pairing agents. The stable isotope L-carnitine-d3 (m/z 165 → 103/85) was used in electrospray MS/MS in the multiple reaction monitoring mode with the ion transitions of m/z 162 → 103/85 for detection and quantitation of L-carnitine. Alkaline hydrolysis of short/medium chain (C2 - C15) acyl-L-carnitines in infant formula was analyzed by LC with quadrupole time-of-flight mass spectrometry (QTOF-MS). The majority of short/medium chain acyl-L-carnitines were hydrolyzed to free L-carnitine. The overall standard deviations for L-carnitine in infant formula, follow-up formula and raw materials ranged from 2.1 to 4.0. The overall mean recoveries ranged from 90.2 to 94.2%.

Publication types

  • Validation Study

MeSH terms

  • Carnitine / analysis*
  • Chromatography, Liquid / methods*
  • Infant Food / analysis*
  • Tandem Mass Spectrometry / methods*

Substances

  • Carnitine