Using a low-amplitude RF pulse at echo time (LARFET) for device localization in MRI

Med Biol Eng Comput. 2014 Oct;52(10):885-94. doi: 10.1007/s11517-014-1184-4. Epub 2014 Aug 31.

Abstract

We describe a new method for frequency down-conversion of MR signals acquired with the radio-frequency projections method for device localization. A low-amplitude, off-center RF pulse applied simultaneously with the echo signal is utilized as the reference for frequency down-conversion. Because of the low-amplitude and large offset from the Larmor frequency, the RF pulse minimally interfered with magnetic resonance of protons. We conducted an experiment with the coil placed at different positions to verify this concept. The down-converted signal was transformed into optical signal and transmitted via fiber-optic cable to a receiver unit placed outside the scanner room. The position of the coil could then be determined by the frequency analysis of this down-converted signal and superimposed on previously acquired MR images for comparison. Because of minimal positional errors (≤ 0.8 mm), this new device localization method may be adequate for most interventional MRI applications.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Humans
  • Magnetic Resonance Imaging / instrumentation*
  • Phantoms, Imaging
  • Radio Waves*
  • Time Factors