The aim of the study described here was to quantitatively assess thermal and mechanical effects of therapeutic ultrasound (US) by sonicating a joint-mimicking phantom, made of muscle-equivalent material, using clinical US equipment. The phantom contains two bone disks simulating a deep joint (treated at 1 MHz) and a superficial joint (3 MHz). Thermal probes were inserted in fixed positions. To test the mechanical (cavitational) effects, we used a latex balloon filled with oxygen-loaded nanobubbles; the dimensions of the oxygen-loaded nanobubbles were determined before and after sonication. Significant increases in temperature (up to 17°C) with fixed field using continuous waves were detected both in front of and behind the bones, depending on the US mode (continuous wave vs. pulsed wave) and on the treatment modality (fixed vs. massage). We found no significant differences in mechanical effects. Although limited by the in vitro design (no blood perfusion, no metabolic compensation), the results can be used to guide operators in their choice of the best US treatment modality for a specific joint.
Keywords: Frequency dependence; Heat; Joint-mimicking phantom; Mechanical effect; Quality control; Rehabilitation; Thermal effect; Ultrasonic therapy.