In vivo precision of the GE Lunar iDXA for the measurement of visceral adipose tissue in adults: the influence of body mass index

Eur J Clin Nutr. 2014 Dec;68(12):1365-7. doi: 10.1038/ejcn.2014.213. Epub 2014 Oct 15.

Abstract

CoreScan is a new software for the GE Lunar iDXA, which provides a quantification of visceral adipose tissue (VAT). The objective of this study was to determine the in vivo precision of CoreScan for the measurement of VAT mass in a heterogeneous group of adults. Forty-five adults (aged 34.6 (8.6) years), ranging widely in body mass index (BMI 26.0 (5.2) kg/m(2); 16.7-42.4 kg/m(2)), received two consecutive total body scans with repositioning. The sample was divided into two subgroups based on BMI, normal-weight and overweight/obese, for precision analyses. Subgroup analyses revealed that precision errors (RMSSD:%CV; root mean square standard deviation:% coefficient of variation) for VAT mass were 20.9 g:17.0% in the normal-weight group and 43.7 g:5.4% in overweight/obese groups. Our findings indicate that precision for DXA-VAT mass measurements increases with BMI, but caution should be used with %CV-derived precision error in normal BMI subjects.

MeSH terms

  • Abdominal Fat / diagnostic imaging*
  • Absorptiometry, Photon / methods*
  • Adult
  • Body Mass Index
  • Female
  • Humans
  • Male
  • Obesity / diagnostic imaging*
  • Reproducibility of Results
  • Software*