Simultaneous determination of the endogenous free α-lipoic acid and dihydrolipoic acid in human plasma and erythrocytes by RP-HPLC coupled with electrochemical detector

Methods Mol Biol. 2015:1208:345-60. doi: 10.1007/978-1-4939-1441-8_25.

Abstract

A highly sensitive, precise, and accurate reversed-phase high performance liquid-chromatography/electrochemical detection method for simultaneous determination of the endogenous free α-lipoic acid and dihydrolipoic acid in biological matrices is presented. The two analytes are extracted from samples with acetonitrile-10% m-phosphoric acid solution(aqueous) (50:50 v/v). To determine the total lipoic acid, samples are treated with tris(2-carboxyethyl)phosphine solution in phosphate buffer: pH 2.5 with 85% o-phosphoric acid prior to deproteination. The two analytes are separated on a C18 (150 × 4.6 mm, 5 μm) analytical column using acetonitrile-50 mM phosphate buffer: pH 2.5 with 85% o-phosphoric acid (35:65 v/v) as the isocratic mobile phase pumped at a flow rate of 2.0 ml/min at the column oven temperature of 35 °C. The column eluents are monitored at a potential of 0.9 V. These analytes are efficiently resolved in <7 min.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Calibration
  • Chromatography, High Pressure Liquid / methods*
  • Chromatography, Reverse-Phase / methods*
  • Electrochemical Techniques / instrumentation*
  • Erythrocytes / metabolism*
  • Humans
  • Hydrodynamics
  • Reference Standards
  • Reproducibility of Results
  • Solutions
  • Solvents
  • Specimen Handling
  • Thioctic Acid / analogs & derivatives*
  • Thioctic Acid / blood*

Substances

  • Solutions
  • Solvents
  • Thioctic Acid
  • dihydrolipoic acid