Comparison of silicone and polyurethane catheters for the catheterization of small vessels in mice

Lab Anim (NY). 2014 Nov;43(11):397-403. doi: 10.1038/laban.570.

Abstract

Vascular catheterization is increasingly carried out in laboratory mice, but the long-term patency of catheters implanted in mice is limited owing to their small size. The authors compared mice that were catheterized in their right common carotid artery either with a silicone catheter with a polyurethane tip or with a 100% polyurethane catheter to determine which catheter type was better suited for long-term studies in mice. The catheters were inspected daily and were flushed if blood was visible in the catheters; if no blood was visible, they were flushed every 3 d. Silicone catheters were patent for a shorter period of time than polyurethane catheters, which were patent for a median of 6 d and up to 25 d. The authors identify the principal causes of catheter failure and discuss how they can be prevented.

MeSH terms

  • Animals
  • Equipment Failure
  • Male
  • Materials Testing
  • Mice
  • Mice, Inbred Strains
  • Polyurethanes / adverse effects*
  • Random Allocation
  • Silicones / adverse effects*
  • Thrombosis / etiology
  • Vascular Access Devices / adverse effects*