Aortic hemodynamics after thoracic endovascular aortic repair, with particular attention to the bird-beak configuration

J Endovasc Ther. 2014 Dec;21(6):791-802. doi: 10.1583/14-4778MR.1.

Abstract

Purpose: To quantitatively evaluate the impact of thoracic endovascular aortic repair (TEVAR) on aortic hemodynamics, focusing on the implications of a bird-beak configuration.

Methods: Pre- and postoperative CTA images from a patient treated with TEVAR for post-dissecting thoracic aortic aneurysm were used to evaluate the anatomical changes induced by the stent-graft and to generate the computational network essential for computational fluid dynamics (CFD) analysis. These analyses focused on the bird-beak configuration, flow distribution into the supra-aortic branches, and narrowing of the distal descending thoracic aorta. Three different CFD analyses (A: preoperative lumen, B: postoperative lumen, and C: postoperative lumen computed without stenosis) were compared at 3 time points during the cardiac cycle (maximum acceleration of blood flow, systolic peak, and maximum deceleration of blood flow).

Results: Postoperatively, disturbance of flow was reduced at the bird-beak location due to boundary conditions and change of geometry after TEVAR. Stent-graft protrusion with partial coverage of the origin of the left subclavian artery produced a disturbance of flow in this vessel. Strong velocity increase and flow disturbance were found at the aortic narrowing in the descending thoracic aorta when comparing B and C, while no effect was seen on aortic arch hemodynamics.

Conclusion: CFD may help physicians to understand aortic hemodynamic changes after TEVAR, including the change in aortic arch geometry, the effects of a bird-beak configuration, the supra-aortic flow distribution, and the aortic true lumen dynamics. This study is the first step in establishing a computational framework that, when completed with patient-specific data, will allow us to study thoracic aortic pathologies and their endovascular management.

Keywords: aorta; biomechanics; bird-beak; computational fluid dynamics; flow disturbance; stent-graft; thoracic endovascular aortic repair; true lumen narrowing.

MeSH terms

  • Aorta, Thoracic / diagnostic imaging
  • Aorta, Thoracic / physiopathology
  • Aorta, Thoracic / surgery*
  • Aortic Aneurysm, Thoracic / diagnosis
  • Aortic Aneurysm, Thoracic / physiopathology
  • Aortic Aneurysm, Thoracic / surgery*
  • Aortography / methods
  • Blood Vessel Prosthesis
  • Blood Vessel Prosthesis Implantation* / adverse effects
  • Blood Vessel Prosthesis Implantation* / instrumentation
  • Computer Simulation
  • Endovascular Procedures* / adverse effects
  • Endovascular Procedures* / instrumentation
  • Hemodynamics*
  • Humans
  • Male
  • Middle Aged
  • Models, Cardiovascular
  • Numerical Analysis, Computer-Assisted
  • Predictive Value of Tests
  • Prosthesis Design
  • Radiographic Image Interpretation, Computer-Assisted
  • Stents
  • Tomography, X-Ray Computed
  • Treatment Outcome