Background: An abduction brace and abduction exercises are commonly employed after humeral greater tuberosity fracture repair. However, the effects of glenohumeral abduction on the biomechanical strength have seldom been elucidated.
Method: We studied 24 pairs of female fresh frozen porcine shoulders randomly divided into 3 groups. After creation of a greater tuberosity fracture on each shoulder, 3 fixation techniques were used for the 3 groups: double-row suture anchor fixation (DR), suture bridge technique (SB), and 2-screw fixation (TS). This biomechanical study was conducted to compare the forces that create 3- and 5-mm displacements and the ultimate failure load at the simulated shoulder abduction angles of 0° and 45° among the 3 groups.
Results: In the DR group, the mean forces to create 3- and 5-mm displacements and the failure load at 0° were higher than those at 45° (P = .036, P = .012, P = .027). By contrast, in the SB group, the mean forces to create 3- and 5-mm displacements at 45° were greater than those at 0° (P = .012, P = .012). There were no significant differences in the forces to create 3- and 5-mm displacements and construct failure between 0° and 45° in the TS group (P = .575, .327, .478).
Conclusion: The DR group had greatest initial fixation strength at a low abduction angle, whereas the SB group had the highest initial fixation strength at a high abduction angle. The TS group appeared unaffected by the abduction angle.
Keywords: Humeral greater tuberosity fracture; abduction; biomechanics; screw; surgery; suture anchor.
Copyright © 2015 Journal of Shoulder and Elbow Surgery Board of Trustees. Published by Elsevier Inc. All rights reserved.