Real-time 3D curved needle segmentation using combined B-mode and power Doppler ultrasound
- PMID: 25485402
- DOI: 10.1007/978-3-319-10470-6_48
Real-time 3D curved needle segmentation using combined B-mode and power Doppler ultrasound
Abstract
This paper presents a real-time segmentation method for curved needles in biological tissue based on analysis of B-mode and power Doppler images from a tracked 2D ultrasound transducer. Mechanical vibration induced by an external voice coil results in a Doppler response along the needle shaft, which is centered around the needle section in the ultrasound image. First, B-mode image analysis is performed within regions of interest indicated by the Doppler response to create a segmentation of the needle section in the ultrasound image. Next, each needle section is decomposed into a sequence of points and transformed into a global coordinate system using the tracked transducer pose. Finally, the 3D shape is reconstructed from these points. The results of this method differ from manual segmentation by 0.71 ± 0.55 mm in needle tip location and 0.38 ± 0.27 mm along the needle shaft. This method is also fast, taking 5-10 ms to run on a standard PC, and is particularly advantageous in robotic needle steering, which involves thin, curved needles with poor echogenicity.
Similar articles
-
3-D ultrasound-guided robotic needle steering in biological tissue.IEEE Trans Biomed Eng. 2014 Dec;61(12):2899-910. doi: 10.1109/TBME.2014.2334309. Epub 2014 Jul 1. IEEE Trans Biomed Eng. 2014. PMID: 25014948 Free PMC article.
-
3D TRUS guided robot assisted prostate brachytherapy.Med Image Comput Comput Assist Interv. 2005;8(Pt 2):17-24. doi: 10.1007/11566489_3. Med Image Comput Comput Assist Interv. 2005. PMID: 16685938
-
Needle and seed segmentation in intra-operative 3D ultrasound-guided prostate brachytherapy.Ultrasonics. 2006 Dec 22;44 Suppl 1:e331-6. doi: 10.1016/j.ultras.2006.07.003. Epub 2006 Jul 28. Ultrasonics. 2006. PMID: 16908040
-
A review of 3D/2D registration methods for image-guided interventions.Med Image Anal. 2012 Apr;16(3):642-61. doi: 10.1016/j.media.2010.03.005. Epub 2010 Apr 13. Med Image Anal. 2012. PMID: 20452269 Review.
-
Review of Robotic Needle Guide Systems for Percutaneous Intervention.Ann Biomed Eng. 2019 Dec;47(12):2489-2513. doi: 10.1007/s10439-019-02319-9. Epub 2019 Jul 31. Ann Biomed Eng. 2019. PMID: 31372856 Review.
Cited by
-
A 3D multi-modal intelligent intervention system using electromagnetic navigation for real-time positioning and ultrasound images: a prospective randomized controlled trial.Ann Transl Med. 2022 Jun;10(11):625. doi: 10.21037/atm-21-5049. Ann Transl Med. 2022. PMID: 35813320 Free PMC article.
-
Shape Reconstruction Processes for Interventional Application Devices: State of the Art, Progress, and Future Directions.Front Robot AI. 2021 Nov 19;8:758411. doi: 10.3389/frobt.2021.758411. eCollection 2021. Front Robot AI. 2021. PMID: 34869615 Free PMC article. Review.
-
Accurate Needle Localization Using Two-Dimensional Power Doppler and B-Mode Ultrasound Image Analyses: A Feasibility Study.Sensors (Basel). 2018 Oct 16;18(10):3475. doi: 10.3390/s18103475. Sensors (Basel). 2018. PMID: 30332743 Free PMC article.
-
Active Localization and Tracking of Needle and Target in Robotic Image-Guided Intervention Systems.Auton Robots. 2018 Jan;42(1):83-97. doi: 10.1007/s10514-017-9640-2. Epub 2017 Jun 12. Auton Robots. 2018. PMID: 29449761 Free PMC article.
-
Methods for Improving the Curvature of Steerable Needles in Biological Tissue.IEEE Trans Biomed Eng. 2016 Jun;63(6):1167-77. doi: 10.1109/TBME.2015.2484262. Epub 2015 Oct 1. IEEE Trans Biomed Eng. 2016. PMID: 26441438 Free PMC article.