Reliability of the i-STAT for the determination of blood electrolyte (K+, Na+, and CI-) concentrations in cattle

J Vet Intern Med. 2015 Jan;29(1):388-94. doi: 10.1111/jvim.12526.

Abstract

Background: Rapid determination of blood electrolyte concentrations can help determine electrolyte status and delivery of effective volume of electrolyte solutions in field conditions.

Objective: To evaluate reliability of the i-STAT, a point-of-care (POC) device, in measuring blood K(+), Na(+), and CI(-) concentrations in cattle.

Animals: Ninety-eight cattle with various diseases.

Methods: In this prospective study, blood samples collected from the jugular vein were processed for determination of K(+), Na(+), and CI(-) concentrations in blood and plasma using the i-STAT and auto-analyzer (Cobas C501), respectively. Blood and plasma electrolyte data were subjected to student t-test for comparison, the concordance analysis for agreement, accuracy, and precision, the Passing-Bablok regression and the Bland-Altman plot for reliability, and receiver operating characteristics curves for sensitivity (Se) and specificity (Sp).

Results: Plasma concentrations of K(+) (4.39 versus 4.2 mmol/L; P < .0001) and CI(-) (100.30 versus 99.4 mmol/L; P < .04) were greater than their concentrations in blood. Plasma and blood Na(+) concentrations were similar (136.95 versus 136.8 mmol/L). The i-STAT results were highly correlated with the Cobas C501 results (r = 0.970, 0.922, and 0.866 for K(+), Na(+), and CI(-), respectively). Regression equations fitting blood (Y) and plasma (X) concentration did not deviate from the identity line for K(+) (Y = -0.10 + 0.98 × X), Na(+) (Y = X), and CI(-) (Y = 3.04 + 0.96 × X). The mean bias (blood concentration - plasma concentration) was -0.20 for K(+) (P = .03), -0.16 for Na(+) (P = .12), and -0.87 for CI(-) (P = .93). The i-STAT had 76-100% Se and 87.7-100% Sp for assessing electrolyte statuses.

Conclusions and clinical importance: The i-STAT yielded results that were in agreement with the auto-analyzer, with negligible biases in measurement of plasma K(+), Na(+), and CI(-) concentrations. The i-STAT is a reliable POC device and can be used in field condition to assess electrolyte status in cattle.

Keywords: Accuracy; Bovine; Point-of-care device; Precision; Sensitivity; Specificity.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Cattle
  • Cattle Diseases / blood*
  • Cattle Diseases / diagnosis
  • Chlorides / blood*
  • Female
  • Hyperkalemia / blood
  • Hyperkalemia / diagnosis
  • Hyperkalemia / veterinary
  • Hypernatremia / blood
  • Hypernatremia / diagnosis
  • Hypernatremia / veterinary
  • Hypokalemia / blood
  • Hypokalemia / diagnosis
  • Hypokalemia / veterinary
  • Hyponatremia / blood
  • Hyponatremia / diagnosis
  • Hyponatremia / veterinary
  • Point-of-Care Systems*
  • Potassium / blood*
  • Sodium / blood*

Substances

  • Chlorides
  • Sodium
  • Potassium