High-performance liquid chromatographic determination of bromazepam in human plasma

Analyst. 1989 May;114(5):639-41. doi: 10.1039/an9891400639.

Abstract

An assay using high-performance liquid chromatography has been developed for the determination of bromazepam in plasma. After a single-step extraction from basified samples with dichloromethane, using decarboxyloflazepate as an internal standard, samples were analysed using a reversed-phase Nova Pak 5-microns column with a mobile phase of methanol - phosphate buffer (60 + 40) adjusted to pH 7.6. The drugs were detected at 239 nm and the limit of detection was found to be 3 micrograms l-1 for bromazepam. The method is simple, rapid and sensitive and permits bromazepam levels in clinical and pharmacokinetic studies to be monitored.

MeSH terms

  • Anti-Anxiety Agents / blood*
  • Bromazepam / blood*
  • Chemical Phenomena
  • Chemistry
  • Chromatography, High Pressure Liquid
  • Humans
  • Spectrophotometry, Ultraviolet

Substances

  • Anti-Anxiety Agents
  • Bromazepam