Background: External thermal support is critical for preterm or ill infants due to altered thermoregulation. Incubators are the gold standard for long-term support and have been adopted successfully in many countries. Alternatives such as radiant warmers, blankets and others are often used as standard of care (SoC) in resource-limited settings when infants are otherwise not in Kangaroo Mother Care (KMC).
Methods: In this pilot study, we evaluate the feasibility of a conductive thermal mattress (CTM) using phase change materials as a low-cost warmer. We conducted a prospective multicentre open-label randomised controlled trial to determine non-inferiority of this CTM to SoC warming practices in low birthweight infants. The primary outcome was maintenance of axillary temperature.
Results: We equally randomised 160 infants to CTM or SoC. The latter cohort continued to receive warmth by radiant warmers (n=48), blankets (n=18), warmed cradles (n=7) or KMC (n=7) before, during and subsequent to the study. CTM was deemed non-inferior since warmed babies had higher axillary temperature compared with SoC (mean increase 0.11±0.03°C SEM; p<0.001). Post hoc comparison to radiant warmers alone showed that CTM led to a higher axillary temperature (mean increase by 0.14±0.03°C SEM; p<0.001).
Conclusions: Short-term use of CTM compared with radiant warmers and other modes of warming is non-inferior to SoC and efficacious in maintaining body temperature. No adverse effects were reported. An extended multinational trial, preferably one that demonstrates longer-term thermoregulation, is warranted.
Trial registration number: Clinical Trials Registry of India (CTRI/2010/091/002916 and CTRI/2011/04/001696).
Keywords: Intensive Care; Neonatology; Technology; Temp Regulation.
Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://group.bmj.com/group/rights-licensing/permissions.