Increase of ventricular interval during atrial fibrillation by atrioventricular node vagal stimulation: chronic clinical atrioventricular-nodal stimulation download study

Circ Arrhythm Electrophysiol. 2015 Jun;8(3):562-8. doi: 10.1161/CIRCEP.114.002588. Epub 2015 Apr 15.

Abstract

Background: Patients with a high ventricular rate during atrial fibrillation (AF) are at increased risk of receiving inappropriate implantable cardioverter defibrillator shocks. The objective was to demonstrate the feasibility of high frequency atrioventricular-nodal stimulation (AVNS) to reduce the ventricular rate during AF to prevent inappropriate implantable cardioverter defibrillator shocks.

Methods and results: Patients with a new atrial lead placement as part of a cardiac resynchronization therapy and defibrillator implant and a history of paroxysmal or persistent AF were eligible. If proper atrial lead position was confirmed, AVNS software was uploaded to the cardiac resynchronization therapy device, tested, and optimized. AVNS was delivered via a right atrial pacing lead positioned in the posterior right atrium. Software allowed initiation of high frequency bursts triggered on rapidly conducted AF. Importantly, the efficacy was evaluated during spontaneous AF episodes between 1 and 6 months after implant. Forty-four patients were enrolled in 4 centers. Successful atrial lead placement occurred in 74%. Median implant time of the AVNS lead was 37 minutes. In 26 (81%) patients, manual AVNS tests increased the ventricular interval by >25%. Between 1 and 6 months, automatic AVNS activations occurred in 4 patients with rapidly conducted AF, and in 3 patients, AVNS slowed the ventricular rate out of the implantable cardioverter defibrillator shock zone. No adverse events were associated with the AVNS software.

Conclusions: The present study demonstrated the feasibility of implementation of AVNS in a cardiac resynchronization therapy and defibrillator system. AVNS increased ventricular interval >25% in 81% of patients. AVNS did not influence the safety profile of the cardiac resynchronization therapy and defibrillator system.

Clinical trial registration: clinicaltrials.gov; Unique Identifier: NCT01095952.

Keywords: atrial fibrillation; shock; vagal stimulation.

Publication types

  • Clinical Trial
  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Algorithms
  • Atrial Fibrillation / diagnosis
  • Atrial Fibrillation / physiopathology
  • Atrial Fibrillation / therapy*
  • Atrioventricular Node / innervation*
  • Cardiac Resynchronization Therapy / methods*
  • Cardiac Resynchronization Therapy Devices
  • Death, Sudden, Cardiac / prevention & control*
  • Defibrillators, Implantable
  • Electric Countershock / adverse effects
  • Electric Countershock / instrumentation
  • Electric Countershock / methods*
  • Electrophysiologic Techniques, Cardiac
  • Feasibility Studies
  • Female
  • Heart Ventricles / physiopathology*
  • Humans
  • Italy
  • Male
  • Prosthesis Design
  • Prosthesis Failure
  • Software Design
  • Time Factors
  • Treatment Outcome
  • Vagus Nerve Stimulation / instrumentation
  • Vagus Nerve Stimulation / methods*

Associated data

  • ClinicalTrials.gov/NCT01095952