Miniature forward-viewing common-path OCT probe for imaging the renal pelvis

Biomed Opt Express. 2015 Mar 6;6(4):1164-71. doi: 10.1364/BOE.6.001164. eCollection 2015 Apr 1.

Abstract

We demonstrate an ultrathin flexible cone-scanning forward-viewing OCT probe which can fit through the working channel of a flexible ureteroscope for renal pelvis imaging. The probe is fabricated by splicing a 200 µm section of core-less fiber and a 150 µm section of gradient-index (GRIN) fiber to the end of a single mode (SM) fiber. The probe is designed for common-path OCT imaging where the back-reflection of the GRIN fiber/air interface is used as the reference signal. Optimum sensitivity was achieved with a 2 degree polished probe tip. A correlation algorithm was used to correct image distortion caused by non-uniform rotation of the probe. The probe is demonstrated by imaging human skin in vivo and porcine renal pelvis ex vivo and is suitable for imaging the renal pelvis in vivo for cancer staging.

Keywords: (110.4500) Optical coherence tomography; (170.2150) Endoscopic imaging; (170.3890) Medical optics instrumentation.